Skip to content

VECTRA®, VASCULAR ACCESS GRAFT 10002-5040-001Recalls & reported problems

Thoratec Corporation · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00813024011323), received between Mar 2, 2022 and Oct 4, 2022 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Mechanical Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Mechanical Problem2
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Hemorrhage/Blood Loss/Bleeding1
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1
  • Thrombosis/Thrombus1
  • Unspecified Tissue Injury1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 4, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD A VECTRA VASCULAR ACCESS GRAFT IMPLANTED ON (B)(6) 2022. INITIAL PUNCTURE AND DIALYSIS WAS PERFORMED ON (B)(6) 2022. AFTER DIALYSIS, THE NEED WAS REMOVED, WITHOUT RESISTANCE, AND GAUZE WAS APPLIED TO STOP THE BLEEDING. WHEN THE GAUZE WAS REMOVED, A THREAD-LIKE SUBSTANCE CAME OUT OF THE PUNCTURE WOUND. THE SUBSTANCE WAS CUT AND THE ARTIFICIAL BLOOD VESSEL WAS STILL IN USE.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Unspecified Tissue Injury

    MAUDE report 15538806 ↗
  • InjuryAug 30, 2022Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED A THROMBUS ATTACHED TO A FREQUENT PUNCTURE SITE. ECHOCARDIOGRAPHY SHOWED LINEAR MATERIAL IN THE LUMEN OF THE GRAFT. THE THROMBUS WAS REMOVED WITH BIOPSY FORCEPS AND A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY WAS PERFORMED.

    Patient problems: Thrombosis/Thrombus

    MAUDE report 15321554 ↗
  • MalfunctionMar 2, 2022Mechanical Problem; Mechanical Problem

    WHEN THE VECTRA GRAFT WAS IMPLANTED UNDER THE SKIN AND PULLED NEAR THE TOP OF THE LOOP, A CRACK APPEARED IN THE OUTER LAYER, SO IT WAS DISCONTINUED AND REMOVED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 13652392 ↗

Other Prosthesis, Vascular Graft, Of Less Then 6mm Diameter devices

All Prosthesis, Vascular Graft, Of Less Then 6mm Diameter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page