TightRail Mini Rotating Dilator Sheath 540-011Recalls & reported problems
Spectranetics Corporation · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00813132021627), received between Feb 12, 2020 and Oct 23, 2023 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Mechanical Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Mechanical Problem1
Patient problems
- Great Vessel Perforation1
- Iatrogenic Source1
- No Known Impact Or Consequence To Patient1
- Pericardial Effusion1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryOct 23, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A LEAD EXTRACTION PROCEDURE COMMENCED TO REMOVE TWO RIGHT VENTRICULAR (RV) LEADS (ONE ABANDONED AND THE OTHER ONE ACTIVE AND LUMENLESS) FOR AN UPGRADE TO A CRT-D SYSTEM. A RIGHT ATRIAL (RA) LEAD WAS ALSO PRESENT IN THE PATIENT, BUT WAS NOT TARGETED FOR EXTRACTION. A SPECTRANETICS LEAD LOCKING DEVICE (LLD EZ) WAS INSERTED INTO THE ABANDONED RV LEAD TO PROVIDE TRACTION. BEGINNING WITH A SPECTRANETICS 14F GLIDELIGHT LASER SHEATH AND A SPECTRANETICS MEDIUM VISISHEATH DILATOR SHEATH, PROGRESS STALLED IN THE INNOMINATE REGION. THEN, A SPECTRANETICS 11F TIGHTRAIL MINI ROTATING DILATOR SHEATH WAS USED TO PROGRESS SLOWLY INTO THE SUPERIOR VENA CAVA (SVC), WITH ADVANCEMENT BEING DIFFICULT DUE TO TIGHT SPACE WITHIN THE VASCULATURE. THE PATIENT''S BLOOD PRESSURE DROPPED AND TRANSESOPHAGEAL ECHOCARDIOGRAPHY (TEE) CONFIRMED THE PRESENCE OF A PERICARDIAL EFFUSION. RESCUE EFFORTS BEGAN, INCLUDING RESCUE BALLOON AND STERNOTOMY. A SMALL (APPROXIMATELY 1-1 1/2 CM) SVC LATERAL WALL PERFORATION WAS DISCOVERED AND REPAIRED. AFTER REPAIR WAS COMPLETED, A 13F TIGHTRAIL WAS USED POST-STERNOTOMY TO ATTEMPT REMOVAL OF THE ABANDONED RV LEAD. HOWEVER, WHEN THE TIGHTRAIL ADVANCED TO THE AREA OF INJURY, THE PERF
Patient problems: Great Vessel Perforation; Pericardial Effusion
MAUDE report 17992728 ↗MalfunctionFeb 12, 2020Mechanical Problem; Mechanical Problem
A LEAD EXTRACTION PROCEDURE COMMENCED TO REMOVE ONE RIGHT VENTRICLE (RV) LEAD DUE TO NON FUNCTION; FIVE OTHER LEADS WERE PRESENT BUT THEY WERE NOT TARGETED FOR EXTRACTION. THE SUBCLAVIAN TO SUPERIOR VENA CAVA (SVC) WAS OCCLUDED MAKING IT DIFFICULT TO ADVANCE A SPECTRANETICS 14F GLIDELIGHT LASER SHEATH. SINCE THE GLIDELIGHT WAS UNABLE TO MOVE THROUGH THE SUBCLAVIAN, A SPECTRANETICS 11F TIGHTRAIL MINI WAS USED TO CLEAR THE OCCLUSION. AFTER ADVANCEMENT WAS MADE WITH THE TIGHTRAIL MINI, A SPECTRANETICS 16F GLIDELIGHT LASER SHEATH AND SPECTRANETICS VISISHEATH WERE INTRODUCED INTO THE SUBCLAVIAN. THE GLIDELIGHT LASER AND VISISHEATH THEN STALLED MAKING LITTLE TO NO PROGRESS. AFTER THE GLIDELIGHT LASER AND VISISHEATH STALLED THE TIGHTRAIL MINI WAS REINTRODUCED BACK INTO THE SUBCLAVIAN. THIS DEVICE MADE SLOW PROGRESS THROUGH THE OCCLUSION. IT WAS DURING THIS ADVANCEMENT THAT THE HANDLE OF THE TIGHTRAIL MINI STOPPED TRIGGERING MAKING THE BLADES UNABLE TO ACTUATE. A NEW SPECTRANETICS 11FR TIGHTRAIL MINI WAS INTRODUCED BACK INTO THE SUBCLAVIAN, THIS NEW TIGHTRAIL MINI MADE SLOW PROGRESS AND THEN STALLED. THE PHYSICIAN THEN ELECTED TO TRY A SPECTRANETICS 13F TIGHTRAIL SUB-C DEVICE FOLLOWED BY A
Patient problems: Iatrogenic Source; No Known Impact Or Consequence To Patient
MAUDE report 9700048 ↗
Other Dilator, Vessel, For Percutaneous Catheterization devices
- VersaCross Connect LAAC Access Solution VXAK0007Baylis Médicale Cie Inc · Dilator, Vessel, For Percutaneous Catheterization3 deaths108 reports
- TandemLife 5740-2417Cardiacassist, Inc. · Dilator, Vessel, For Percutaneous Catheterization2 recalls0 reports
- TightRail Rotating Dilator Sheath 545-513Spectranetics Corporation · Dilator, Vessel, For Percutaneous Catheterization20 deaths74 reports
- VersaCross Connect LAAC Access Solution VXAK0003Baylis Médicale Cie Inc · Dilator, Vessel, For Percutaneous Catheterization2 deaths70 reports
- VersaCross Connect LAAC Access Solution VXAK0007Baylis Médicale Cie Inc · Dilator, Vessel, For Percutaneous Catheterization2 deaths54 reports
- AVANTI 504656XCordis Corporation · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION1 recall2 reports
- Mini Stick MAX H965457510Angiodynamics, Inc. · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION1 recall1 report
- Abiomed® 00884450732311Merit Medical Systems, Inc. · Dilator, Vessel, For Percutaneous Catheterization1 recall1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page