AngioSculpt OTW PTA Scoring Balloon Catheter with HydroCross Coating 2237-40100Recalls & reported problems
The Spectranetics Corporation · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00813132022822), received between Jan 21, 2021 and Jan 21, 2021 — classified as 1 malfunction.
The most frequently reported specific device problem is “Material Rupture” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Rupture1
Patient problems
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 21, 2021Material Rupture; Material Rupture
THE ANGIOSCULPT DEVICE WAS USED TO TREAT A MODERATELY CALCIFIED MID SFA. DURING INFLATION AT 8 ATM, A PINHOLE LEAK WAS DETECTED SINCE CONTRAST COULD BE SEEN ON THE ANGIO. THE PROCEDURE WAS COMPLETED WITH A NEW DEVICE. NO PATIENT INJURY REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11211150 ↗
Other Catheter, Percutaneous, Cutting/Scoring devices
- AngioSculpt OTW PTA Scoring Balloon Catheter 2092-6040The Spectranetics Corporation · Catheter, Percutaneous, Cutting/Scoring41 reports
- AngioSculpt OTW PTA Scoring Balloon Catheter 2076-5040The Spectranetics Corporation · Catheter, Percutaneous, Cutting/Scoring31 reports
- AngioSculpt OTW PTA Scoring Balloon Catheter 2092-5040The Spectranetics Corporation · Catheter, Percutaneous, Cutting/Scoring28 reports
- 2cm Peripheral Cutting Balloon® M001PCB8020900Boston Scientific Corporation · Catheter, percutaneous, cutting/scoring18 reports
- AngioSculpt OTW PTA Scoring Balloon Catheter 2076-6040The Spectranetics Corporation · Catheter, Percutaneous, Cutting/Scoring17 reports
- AngioSculpt OTW PTA Scoring Balloon Catheter 2105-6040The Spectranetics Corporation · Catheter, Percutaneous, Cutting/Scoring17 reports
- ULTRASCORE US14150530Bard Peripheral Vascular, Inc. · Catheter, percutaneous, cutting/scoring15 reports
- ULTRASCORE US14150615Bard Peripheral Vascular, Inc. · Catheter, percutaneous, cutting/scoring14 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page