primePICC PFM2CT4DPRecalls & reported problems

Pfm Medical Incorporated · 4F x 50cm Dual Lumen CT PICC Nursing Kit · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00813623010185), received between Jan 24, 2022 and Jan 24, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Crack” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Crack1
  • Device Damaged by Another Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 24, 2022Crack; Device Damaged by Another Device; Crack; Device Damaged by Another Device

    PICC WAS PLACED WITHOUT ISSUE. A FEW DAYS LATER THE LINE WAS LEAKING AROUND A CAP BUT THOUGHT TO BE A CAP ISSUE NOT A LINE ISSUE, THEN THE ACTUAL LUMEN OF THE CATHETER WAS NOTED TO BE CRACKED 4-5 DAYS AFTER PLACEMENT. THE CAPS WE USE ARE FROM ICU MEDICAL, INC. MICROCLAVE CLEAR NEUTRAL CONNECTER. PATIENT WAS FINE; NEW PERIPHERAL LINE WAS PLACED WITHOUT ISSUE AFTER PFM PICC LINE DEVELOPED CRACKED LUER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13336932 ↗

Other Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days devices

All Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days devices →

Sources and updates