Skip to content

LifeSPARC 5800-0000Recalls & reported problems

Cardiacassist, Inc. · LifeSPARC blood pump kit with accessories · Class 1

All models of LifeSPARC (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00814112020593), received between Jul 3, 2023 and Nov 3, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Pumping Stopped” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Pumping Stopped2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 3, 2023Pumping Stopped; Pumping Stopped

    LIVANOVA RECEIVED A REPORT THAT A LIFESPARC PUMP STOPPED AND WAS UNABLE TO BE RESTARTED DURING SUPPORT. THE USER ATTEMPTED TO PRIME THE PUMP WITHOUT A LIVANOVA REPRESENTATIVE PRESENT AND DESCRIBED PUTTING AN UNSPECIFIED AMOUNT OF SALINE INTO THE CIRCUIT AND PLACING BOTH ENDS OF THE CIRCUIT INTO A BASIN TO ATTEMPT A BASIN PRIME. THE PUMP WAS STARTED AND INITIALLY RAN, BUT USER REPORTED THAT IT STOPPED SHORTLY AFTER AND COULD NOT BE RESTARTED. THE USER REPORTEDLY ATTEMPTED A RESTART OF THE LIFESPARC CONTROLLER, PLUGGING IN THE DRIVE LINE SEVERAL TIMES WITH NO IMPROVEMENT. THE USER INDICATED THAT THEY POTENTIALLY HAD AN "AIR LOCK" IN THE PUMP HEAD OR THAT THE DRIVE LINE PLUG POTENTIALLY GOT WET BUT WERE UNABLE TO SPECIFY FURTHER. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18069058 ↗
  • MalfunctionJul 3, 2023Pumping Stopped; Pumping Stopped

    LIVANOVA RECEIVED A REPORT THAT A LIFESPARC PUMP BEGAN MAKING NOISES AND STOPPED DURING SUPPORT. THE NOISY PUMP WAS NOTICED BY A FAMILY MEMBER OF THE PATIENT AND WHEN INSPECTED BY A NURSE, IT WAS OBSERVED THAT THE PUMP HAD COMPLETELY STOPPED AND THE CONTROLLER BEGAN ALARMING. THE PATIENT WAS ON SUPPORT FOR 10 DAYS AT THE TIME OF THE PUMP STOP. THE ENTIRE CIRCUIT WAS SWAPPED OUT AND REPLACED AND THE ISSUE WAS RESOLVED. NO HARM TO THE PATIENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17252959 ↗

Other Accessory Equipment, Cardiopulmonary Bypass devices

All Accessory Equipment, Cardiopulmonary Bypass devices →

Sources and updates