Skip to content

LifeSPARCRecalls & reported problems across all models

Cardiacassist, Inc. · 5 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name at least one LifeSPARC model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
16
Deaths reported
2
Injuries reported
1
Recalls
4

We found 5 models of LifeSPARC by Cardiacassist, Inc. in the FDA's device registry (GUDID). Together they are named in 16 adverse event reports in the FDA's MAUDE database, received between Jan 31, 2022 and Jan 29, 2025 — classified as 2 deaths, 1 injury, 13 malfunctions.

The most frequently reported specific device problem is “Pumping Stopped” (6 reports).

4 FDA recall records name at least one of these models by its device identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 1 reportsDec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

LifeSPARC recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Jun 30, 2023Open, Classified

    Epoxy used during manufacturing of the LifeSPARC Pump may have uncured epoxy. if undetectable curing may result in added risk of early pump failure,

    Z-2465-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Dec 5, 2022Terminated

    A software update (v1.1.5) has been developed to address the issue of Critical Failure which can occur on the LifeSPARC Controller. The Critical Failure was first addressed in the firm's recall initiated July 21, 2022. During the Critical Failure, the software freezes or crashes and the screen does not display data.

    Z-0926-2023 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗

  3. Jul 21, 2022Terminated

    Critical Failure of the LifeSPARC controller when the software freezes or crashes and the screen does not display data.

    Z-1763-2022 · Root cause: Under Investigation by firm · names 2 of these models · Recall details · FDA record ↗

  4. Oct 14, 2021Open, Classified

    Top label of the outer package of the collection kit Item 5820-3118 showed expiration date of 2022/05/01 (correct expiration date) while side label showed expiration date of 2021/05/01 (incorrect expiration date)

    Z-0311-2022 · Root cause: Error in labeling · Recall details · FDA record ↗

Most reported LifeSPARC problems

Device problems

  • Pumping Stopped6
  • Infusion or Flow Problem4
  • Adverse Event Without Identified Device or Use Problem2
  • Backflow2
  • Battery Problem1
  • Crack1
  • Detachment of Device or Device Component1
  • Insufficient Flow or Under Infusion1

Patient problems

  • No Clinical Signs, Symptoms or Conditions13
  • Reduced Blood Flow2
  • Cardiac Arrest1

LifeSPARC models

5 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent LifeSPARC adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page