Skip to content

Zyno BX-70071-DF-120Recalls & reported problems

Zyno Medical LLC · BX Administration Primary Set: 105" 1 NF Y Site Below Pump, 1.2 Filter, Anti-Free Flow Clamp. BX Primary Administration Sets are single-use IV administration sets for use with Z-800F, Z-800W and Z-800WF Zyno Medical Infusion pumps. All BX Primary sets fluid paths are sterile, non-Pyrogenic, are not made with natural rubber latex, and are not made with DEHP. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00814371020556), received between May 20, 2024 and May 20, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Contamination” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Contamination1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 20, 2024Contamination; Contamination

    ON 05/02/2024, ZYNO RECEIVED A REPORT FROM A CUSTOMER WHOM HAD PURCHASED AX-80075 LOT 2302002 AND BX-70071-DF LOT 2309023 FROM INFUSYSTEM, THEY STATED THEY FOUND HAIRS INSIDE THE TUBING BAGS. THE CUSTOMER WAS GOING TO BE RETURNING THE REMAINING STOCK OF THESE TWO LOTS TO INFUSYSTEM. ZYNO MEDICAL HAS QUARANTINED ANY OF THE REMAINING STOCK OF THESE LOTS. ZYNO MEDICAL, IS ATTEMPTING TO OBTAIN THE ACTUAL SAMPLES TO SEND IN FOR AN INVESTIGATION. NO PATIENT WAS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19355100 ↗

Other Set, Administration, Intravascular devices

All Set, Administration, Intravascular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page