Penumbra Coil 400 4006C0303-ARecalls & reported problems
Penumbra, Inc. · Coil 400 Extra Soft, 3x3 · Class 2
All models of Penumbra Coil 400 (120)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00814548018157), received between Oct 20, 2023 and Oct 20, 2023 — classified as 1 malfunction.
The most frequently reported specific device problem is “Migration” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Migration1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 20, 2023Migration; Migration
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE ANTERIOR COMMUNICATING ARTERY USING PENUMBRA COIL 400S (PC400S), PENUMBRA COIL 400 DETACHMENT HANDLES (HANDLE), AND A PX SLIM DELIVERY MICROCATHETER (PX SLIM). IT WAS NOTED THAT THE PATIENT''S ANATOMY WAS TORTUOUS. DURING THE PROCEDURE, THE PHYSICIAN IMPLANTED TWO PC400S USING THE HANDLE. AFTER ADVANCING THE THIRD PC400 INTO THE TARGET LOCATION AND MAKING SEVERAL ATTEMPTS TO DETACH THE PC400 USING THE HANDLE, THE COIL FAILED TO DETACH. WHILE RETRACTING THE PUSHER ASSEMBLY OF THE PC400, THE PC400 WAS PULLED BACK INTO THE PX SLIM AND SUBSEQUENTLY UNINTENTIONALLY DETACHED INSIDE THE PX SLIM. THE PX SLIM CONTAINING THE DETACHED PC400 WAS REMOVED FROM THE PATIENT, AND THE DEVICES WERE NO LONGER USED IN THE PROCEDURE. THE PHYSICIAN THEN ADVANCED ANOTHER PC400 INTO THE TARGET LOCATION USING A NEW PX SLIM AND ATTEMPTED TO DETACH THE COIL USING A SECOND HANDLE. AFTER SEVERAL ATTEMPTS, THE PHYSICIAN WAS ABLE SUCCESSFULLY DETACH THE PC400. IT WAS REPORTED THAT ONE LOOP OF THE PC400 WAS EXTRUDING FROM THE ANEURYSM; HOWEVER, THE PHYSICIAN DOES NOT BELIEVE IT WOULD RESULT IN ANY CLINICAL CONSEQUENCE AND NO FURTHER ACTION WAS TAKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17972055 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page