Penumbra, Inc. · Coil 400 Extra Soft, 4x6 · Class 2

All models of Penumbra Coil 400 (120)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00814548018218), received between Jan 29, 2021 and Oct 20, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Physical Resistance/Sticking” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Physical Resistance/Sticking1
  • Premature Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 20, 2023Premature Separation; Premature Separation

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE ANTERIOR COMMUNICATING ARTERY USING PENUMBRA COIL 400S (PC400S), PENUMBRA COIL 400 DETACHMENT HANDLES (HANDLE), AND A PX SLIM DELIVERY MICROCATHETER (PX SLIM). IT WAS NOTED THAT THE PATIENT''S ANATOMY WAS TORTUOUS. DURING THE PROCEDURE, THE PHYSICIAN IMPLANTED TWO PC400S USING THE HANDLE. AFTER ADVANCING THE THIRD PC400 INTO THE TARGET LOCATION AND MAKING SEVERAL ATTEMPTS TO DETACH THE PC400 USING THE HANDLE, THE COIL FAILED TO DETACH. WHILE RETRACTING THE PUSHER ASSEMBLY OF THE PC400, THE PC400 WAS PULLED BACK INTO THE PX SLIM AND SUBSEQUENTLY UNINTENTIONALLY DETACHED INSIDE THE PX SLIM. THE PX SLIM CONTAINING THE DETACHED PC400 WAS REMOVED FROM THE PATIENT, AND THE DEVICES WERE NO LONGER USED IN THE PROCEDURE. THE PHYSICIAN THEN ADVANCED ANOTHER PC400 INTO THE TARGET LOCATION USING A NEW PX SLIM AND ATTEMPTED TO DETACH THE COIL USING A SECOND HANDLE. AFTER SEVERAL ATTEMPTS, THE PHYSICIAN WAS ABLE SUCCESSFULLY DETACH THE PC400. IT WAS REPORTED THAT ONE LOOP OF THE PC400 WAS EXTRUDING FROM THE ANEURYSM; HOWEVER, THE PHYSICIAN DOES NOT BELIEVE IT WOULD RESULT IN ANY CLINICAL CONSEQUENCE AND NO FURTHER ACTION WAS TAKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17972054 ↗
  • MalfunctionJan 29, 2021Physical Resistance/Sticking; Physical Resistance/Sticking

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE INTERNAL CAROTID ARTERY (ICA) AND EXTERNAL CAROTID ARTERY (ECA) USING PENUMBRA COILS 400 (PC400S) AND A PX SLIM DELIVERY MICROCATHETER (PX SLIM). DURING THE PROCEDURE, THE PHYSICIAN IMPLANTED THREE POD COILS IN THE TARGET VESSEL. SUBSEQUENTLY, WHILE ATTEMPTING TO INSERT THE FOURTH COIL, A PC400, INTO THE PX SLIM, THE PHYSICIAN EXPERIENCED RESISTANCE. THEREFORE, THE PC400 WAS REMOVED. THE PROCEDURE WAS COMPLETED USING PENUMBRA COILS 400 (PC400S), POD COILS, AND THE SAME PX SLIM. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 11254281 ↗

Other DEVICE, NEUROVASCULAR EMBOLIZATION devices

All DEVICE, NEUROVASCULAR EMBOLIZATION devices →