Penumbra Coil 400 4004C0513-ARecalls & reported problems

Penumbra, Inc. · Coil 400 Soft, 5x13 · Class 2

All models of Penumbra Coil 400 (120)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00814548018461), received between Jun 27, 2025 and Jun 27, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture1
  • Premature Separation1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 27, 2025Fracture; Premature Separation; Fracture; Premature Separation

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE ARTERIOVENOUS (AV) FISTULA ARTERY USING PENUMBRA COILS 400 (PC400), A PENUMBRA COIL 400 DETACHMENT HANDLE (HANDLE), AN ACCESS25 DELIVERY MICROCATHETER (ACCESS25), AND A GUIDEWIRE. DURING THE PROCEDURE, THE PHYSICIAN ADVANCED A PC400 TO THE TARGET LOCATION USING THE MICROCATHETER. THE PHYSICIAN DETERMINED THE PC400 WAS UNDERSIZED AND DECIDED TO RESHEATH THE PC400. THE PC400 BECAME KNOTTED AND WHILE RETRACTING, THE EMBOLIZATION COIL DETACHED. MANUAL SUCTION AND DUAL SNARES WERE USED TO ATTEMPT TO REMOVE THE EMBOLIZATION COIL WITHOUT SUCCESS. THE PHYSICIAN DECIDED TO LEAVE THE EMBOLIZATION COIL IN THE INTERNAL ILIAC VEIN. IT WAS REPORTED THAT THE EMBOLIZATION COIL MIGRATED TO THE LUNG. NO INTERVENTION WAS PERFORMED TO REMOVE THE EMBOLIZATION COIL. THERE WERE NO ADDITIONAL ADVERSE EVENTS CAUSED BY LEAVING THE DETACHED COIL IN LUNG. THE PHYSICIAN ADVANCED ANOTHER PC400 INTO THE TARGET VESSEL AND ATTEMPTED TO DETACH THE PC400 USING A HANDLE. THE PUSHER ASSEMBLY OF THE PC400 HAD SEPARATED, BUT THE EMBOLIZATION COIL DID NOT DETACH. THE PHYSICIAN THEN PUSHED THE PC400 INTO THE TARGET LOCATION USING THE PUSHER ASSEMBLY AND IT EV

    Patient problems: Foreign Body In Patient

    MAUDE report 22344509 ↗