Penumbra, Inc. · Ruby Soft, 20x60 · Class 2

All models of RUBY Coil (90)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00814548019123), received between Jan 25, 2019 and Apr 15, 2024 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Failure to Advance” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Advance2
  • Premature Separation2
  • Difficult or Delayed Separation1
  • Physical Resistance/Sticking1
  • Separation Failure1

Patient problems

  • No Consequences Or Impact To Patient3
  • No Clinical Signs, Symptoms or Conditions2
  • Foreign Body In Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 15, 2024Difficult or Delayed Separation; Premature Separation; Difficult or Delayed Separation; Premature Separation

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE GONADAL VEIN USING RUBY COILS AND A PX SLIM DELIVERY MICROCATHETER (PX SLIM). DURING THE PROCEDURE, THE PHYSICIAN SUCCESSFULLY IMPLANTED NINE RUBY COILS INTO THE TARGET LOCATION. AFTER ADVANCING THE NEXT RUBY COIL INTO THE TARGET LOCATION, THE PHYSICIAN MADE AN ATTEMPT TO DETACH THE RUBY COIL MANUALLY WITHOUT SUCCESS. THE PHYSICIAN THEN ATTEMPTED TO RE-POSITION THE COIL AND, WHILE INSERTING THE LAST LOOP OF THE RUBY COIL, THE PHYSICIAN NOTICED UNDER FLUOROSCOPY THAT THE RUBY COIL HAD UNINTENTIONALLY DETACHED INSIDE THE VESSEL. THEREFORE, THE RUBY COIL WAS REMOVED USING A SNARE DEVICE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED AT THIS POINT AND NO ADDITIONAL COILS WERE REQUIRED.

    Patient problems: Foreign Body In Patient

    MAUDE report 19106116 ↗
  • MalfunctionMay 25, 2021Separation Failure; Separation Failure

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN AN ENDOLEAK USING A RUBY COIL DETACHMENT HANDLE (HANDLE), RUBY COILS, AND A LANTERN DELIVERY MICROCATHETER (LANTERN). DURING THE PROCEDURE, THE PHYSICIAN ADVANCED A RUBY COIL THROUGH THE LANTERN AND ATTEMPTED TO DETACH THE RUBY COIL WITH THE HANDLE. WHEN THE RUBY COIL WOULD NOT DETACH, THE HOSPITAL TECHNOLOGIST ATTEMPTED TO MANUALLY DETACH THE RUBY COIL BY MOVING THE MANDREL LATERALLY. WITH SIGNIFICANT FORCE, THE RUBY COIL WAS DETACHED SUCCESSFULLY. THE PROCEDURE WAS COMPLETED USING NEW RUBY COILS AND THE SAME HANDLE. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11881050 ↗
  • MalfunctionFeb 24, 2021Failure to Advance; Failure to Advance

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE LEFT SUBCLAVIAN ARTERY USING RUBY COILS, A LANTERN DELIVERY MICROCATHETER (LANTERN), AND A SHEATH. DURING THE PROCEDURE, THE FIRST RUBY COIL WOULD NOT ADVANCE PAST THE MIDDLE OF THE LANTERN. THEREFORE, THE RUBY COIL WAS REMOVED. SUBSEQUENTLY, TWO RUBY COILS WERE SUCCESSFULLY IMPLANTED INTO THE TARGET LOCATION. NEXT, TWO RUBY COILS WOULD NOT ADVANCE PAST THE MIDDLE OF THE LANTERN. THEREFORE, THESE RUBY COILS WERE REMOVED. THE PHYSICIAN DECIDED TO REMOVE THE LANTERN AND, UPON REMOVAL, THE PHYSICIAN NOTICED THE LANTERN WAS KINKED PROXIMAL TO THE HUB. THEREFORE, THE LANTERN WAS NO LONGER USED IN THE PROCEDURE. THE PHYSICIAN USED A NEW LANTERN TO SUCCESSFULLY IMPLANT ONE RUBY COIL. HOWEVER, THE NEXT RUBY COIL WOULD NOT ADVANCE PAST THE DISTAL END OF THE LANTERN. THEREFORE, THE RUBY COIL WAS REMOVED. THE PROCEDURE WAS COMPLETED USING THE SAME LANTERN, ANOTHER RUBY COIL, AND THE SAME SHEATH. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 11376173 ↗
  • InjuryDec 23, 2019Premature Separation; Premature Separation

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE RIGHT GONADAL VEIN USING RUBY COILS, A NON-PENUMBRA MICROCATHETER, A NON-PENUMBRA DIAGNOSTIC CATHETER, AND A NON-PENUMBRA SHEATH. DURING THE PROCEDURE, SIXTEEN COILS WERE SUCCESSFULLY IMPLANTED IN THE TARGET LOCATION. WHILE ADVANCING A RUBY COIL INTO THE TARGET VESSEL, THE DIAGNOSTIC CATHETER AND SHEATH MIGRATED UNBEKNOWNST TO THE PHYSICIAN. SUBSEQUENTLY, THE MICROCATHETER AND RUBY COIL MIGRATED OUT OF THE TARGET VESSEL. THE PHYSICIAN THEN ATTEMPTED TO RETRACT THE RUBY COIL AND NAVIGATE THE MICROCATHETER AND DIAGNOSTIC CATHETER BACK TO THE TARGET LOCATION; HOWEVER, THE RUBY COIL UNINTENTIONALLY DETACHED. IN AN ATTEMPT TO REMOVE THE DETACHED RUBY COIL, ALL DEVICES WERE WITHDRAWN; HOWEVER, THE PROXIMAL PART OF THE RUBY COIL REMAINED IN THE INFERIOR VENA CAVA (IVC). THEREFORE, THE PHYSICIAN USED A SNARE TO SUCCESSFULLY REMOVE THE RUBY COIL. THE PROCEDURE WAS COMPLETED USING THE SAME CATHETERS AND ADDITIONAL RUBY COIL. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9513329 ↗
  • MalfunctionJan 25, 2019Failure to Advance; Physical Resistance/Sticking; Failure to Advance; Physical Resistance/Sticking

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE THORACIC AORTA USING RUBY COILS. DURING THE PROCEDURE, THE PHYSICIAN SUCCESSFULLY PLACED SEVERAL COILS IN THE TARGET VESSEL USING A LANTERN DELIVERY MICROCATHETER (LANTERN). WHILE ATTEMPTING TO ADVANCE A RUBY COIL, THE PHYSICIAN EXPERIENCED RESISTANCE AND WAS UNABLE TO ADVANCE THE RUBY COIL FROM ITS INITIAL POSITION IN THE INTRODUCER SHEATH; THEREFORE, THE RUBY COIL WAS REMOVED. THE PROCEDURE WAS COMPLETED USING A NEW RUBY COIL AND OTHER COILS WITH THE SAME LANTERN. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8279000 ↗

Other Device, Vascular, For Promoting Embolization devices

All Device, Vascular, For Promoting Embolization devices →