Contour 24 Q7000183Recalls & reported problems
Quality Electrodynamics LLC · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00814848020423), received between Jun 23, 2023 and Jun 23, 2023 — classified as 1 injury.
The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Improper or Incorrect Procedure or Method1
Patient problems
- Partial thickness (Second Degree) Burn1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJun 23, 2023Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
QUALITY ELECTRODYNAMICS RECEIVED A COMPLAINT FORWARDED FROM SIEMENS THAT A CONTOUR 24 RF COIL CAUSED A PATIENT BURN. THE BURN WAS ON BACKSIDE OF RIGHT WRIST AND WAS A DIME - NICKEL SIZED BLISTER. THE PATIENT WAS WEARING SCRUBS PROVIDED BY THE HOSPITAL AND WAS NOT SEDATED. THE PATIENT DID NOT REQUEST THAT THE SCAN BE STOPPED. THE PATIENT CHOSE NOT TO SEEK ADDITIONAL MEDICAL CARE FOR THE BURN. THE CONTOUR COIL WAS BEING USED BY ITSELF (WITH NO OTHER COILS) AND WAS POSITIONED OVER THE PATIENT'S PELVIC REGION. PER THE TECHNOLOGIST'S MEMORY: 1. THERE WERE NO PADS BETWEEN THE PATIENTS LEGS. 2. THERE WAS PADDING BETWEEN THE COIL AND BORE. 3. IT IS UNKNOWN IF THERE WERE PADS BETWEEN THE PATIENT' HIPS AND ARMS. 4. IT IS UNKNOWN IF THERE WERE PADS BETWEEN THE PATIENT'S HAND AND THE COIL, BUT THE TECHNICIAN STATED THAT THE PATIENT'S HAND WAS NOT CONTACTING THE COIL AT THE BEGINNING OF THE SCAN. IT IS BELIEVED THAT THE PATIENT MAY HAVE INADVERTENTLY SHIFTED HER/THE COIL'S POSITION DURING THE SCAN, WHICH COULD HAVE CAUSED THE HARD HOUSING OF THE COIL TO CONTACT HER WRIST RESULTING IN THE BURN. THE COIL WAS RETURNED TO SIEMENS AND SUBSEQUENTLY TO QUALITY ELECTRODYNAMICS AFTER THE INCIDENT.
Patient problems: Partial thickness (Second Degree) Burn
MAUDE report 17191593 ↗
FDA clearance and classification
Cleared through 510(k) K183111 on Dec 7, 2018 (decision: Substantially Equivalent). Submitted by Quality Electrodynamics, LLC as “Contour 24”. The FDA review took 29 days. FDA record ↗
The FDA regulates product code MOS (coil, magnetic resonance, specialty) under 21 CFR 892.1000 as a Class II device. 21 CFR 892.1000 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Contour 24 Q7000183
Has the Contour 24 Q7000183 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00814848020423), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Contour 24 Q7000183?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Jun 23, 2023. That report lists device problem “Improper or Incorrect Procedure or Method” (1); patient problem “Partial thickness (Second Degree) Burn” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Contour 24 Q7000183?
None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Contour 24 Q7000183 FDA approved?
It was cleared through 510(k) K183111 on Dec 7, 2018. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Contour 24 Q7000183, and is it still sold?
The FDA's device database lists Quality Electrodynamics LLC as the company responsible for it (brand Contour 24, model Q7000183). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Coil, Magnetic Resonance, Specialty devices
- NA SIGNA PremierGE Medical Systems, L.L.C. · Coil, magnetic resonance, specialty3 recalls0 reports
- NA SIGNA PremierGE Medical Systems, L.L.C. · Coil, magnetic resonance, specialty3 recalls0 reports
- MAGNETOM Skyra 10849580Siemens Healthcare GmbH · Coil, magnetic resonance, specialty2 recalls2 reports
- MAGNETOM Altea 11410371Siemens Healthcare GmbH · Coil, magnetic resonance, specialty1 recall2 reports
- MAGNETOM Spectra 10837643Siemens Shenzhen Magnetic Resonance Ltd. · Coil, magnetic resonance, specialty1 recall1 report
- Invivo 45353022887Philips Ds North America LLC · Coil, magnetic resonance, specialty1 recall0 reports
- Philips dS Breast 16ch 3.0TPhilips Ds North America LLC · Coil, magnetic resonance, specialty1 recall0 reports
- Philips SENSE Breast CoilPhilips Ds North America LLC · Coil, magnetic resonance, specialty1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page