Skip to content

MAGNETOM Altea 11410371Recalls & reported problems

Siemens Healthcare GmbH · MRI Scanner · Class 2

All models of MAGNETOM Altea (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04056869230726), received between Aug 23, 2024 and Jan 15, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Defective Component” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 1, 2023Class IIOpen, Classified

    In some cases, when lubricating grease is present within the headband of the over-ear headphones, it may cause visible, dot- or lineshaped, fat-isointense artifacts during head examinations only. These potential artifacts depend on the position of the headband, the amount of lubricating grease, as well as the sequence parameters used

    Z-2648-2023 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Defective Component1
  • Improper or Incorrect Procedure or Method1
  • Use of Device Problem1

Patient problems

  • Burn(s)1
  • Full thickness (Third Degree) Burn1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 15, 2025Defective Component; Defective Component

    IT WAS REPORTED TO SIEMENS THAT AN ISSUE WAS DISCOVERED ON THE MAGNETOM ALTEA (CN). IT WAS STATED THAT THE CUSTOMER FOUND THE LOWER PART OF THE BM HEAD/NECK 20 COIL SURFACE HAD BEEN BURNED WITH A HOLE AND THE DIAMETER WAS ABOUT 10 MM. THERE WAS NO INJURY ASSOCIATED WITH THIS ISSUE. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO INVESTIGATE THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21156537 ↗
  • InjuryAug 23, 2024Use of Device Problem; Improper or Incorrect Procedure or Method; Use of Device Problem; Improper or Incorrect Procedure or Method

    IT WAS REPORTED TO SIEMENS THAT AN ADVERSE EVENT OCCURRED WHILE OPERATING THE MAGNETOM ALTEA MR SYSTEM. IT WAS STATED THAT A PATIENT REPORTED A SKIN BURN SEVEN DAYS AFTER AN MR SCAN. THE PATIENT NOTED PAIN AND BLISTERS ON THE INDEX KNUCKLE AND LEFT FOREARM WHEN HE GOT IN THE CAR AFTERWARDS. LATER IT WAS DIAGNOSED THAT THE PATIENT SUFFERED A FULL THICKNESS BURN ON THE LEFT FOREARM AND A PARTIAL THICKNESS BURN ON THE RIGHT INDEX KNUCKLE. BOTH BURNS HAVE A SIZE LIKE A KNUCKLE. IT WAS STATED THAT THE POINTS OF BURN WERE TOUCHING EACH OTHER DURING EXAMINATION.

    Patient problems: Burn(s); Full thickness (Third Degree) Burn

    MAUDE report 20046046 ↗

Other Coil, magnetic resonance, specialty devices

All Coil, magnetic resonance, specialty devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page