MAGNETOM Altea 11410371Recalls & reported problems
Siemens Healthcare GmbH · MRI Scanner · Class 2
All models of MAGNETOM Altea (2)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04056869230726), received between Aug 23, 2024 and Jan 15, 2025 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Defective Component” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Sep 1, 2023Class IIOpen, Classified
In some cases, when lubricating grease is present within the headband of the over-ear headphones, it may cause visible, dot- or lineshaped, fat-isointense artifacts during head examinations only. These potential artifacts depend on the position of the headband, the amount of lubricating grease, as well as the sequence parameters used
Z-2648-2023 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Defective Component1
- Improper or Incorrect Procedure or Method1
- Use of Device Problem1
Patient problems
- Burn(s)1
- Full thickness (Third Degree) Burn1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 15, 2025Defective Component; Defective Component
IT WAS REPORTED TO SIEMENS THAT AN ISSUE WAS DISCOVERED ON THE MAGNETOM ALTEA (CN). IT WAS STATED THAT THE CUSTOMER FOUND THE LOWER PART OF THE BM HEAD/NECK 20 COIL SURFACE HAD BEEN BURNED WITH A HOLE AND THE DIAMETER WAS ABOUT 10 MM. THERE WAS NO INJURY ASSOCIATED WITH THIS ISSUE. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO INVESTIGATE THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21156537 ↗InjuryAug 23, 2024Use of Device Problem; Improper or Incorrect Procedure or Method; Use of Device Problem; Improper or Incorrect Procedure or Method
IT WAS REPORTED TO SIEMENS THAT AN ADVERSE EVENT OCCURRED WHILE OPERATING THE MAGNETOM ALTEA MR SYSTEM. IT WAS STATED THAT A PATIENT REPORTED A SKIN BURN SEVEN DAYS AFTER AN MR SCAN. THE PATIENT NOTED PAIN AND BLISTERS ON THE INDEX KNUCKLE AND LEFT FOREARM WHEN HE GOT IN THE CAR AFTERWARDS. LATER IT WAS DIAGNOSED THAT THE PATIENT SUFFERED A FULL THICKNESS BURN ON THE LEFT FOREARM AND A PARTIAL THICKNESS BURN ON THE RIGHT INDEX KNUCKLE. BOTH BURNS HAVE A SIZE LIKE A KNUCKLE. IT WAS STATED THAT THE POINTS OF BURN WERE TOUCHING EACH OTHER DURING EXAMINATION.
Patient problems: Burn(s); Full thickness (Third Degree) Burn
MAUDE report 20046046 ↗
Other Coil, magnetic resonance, specialty devices
- NA SIGNA PremierGE Medical Systems, L.L.C. · Coil, magnetic resonance, specialty3 recalls0 reports
- NA SIGNA PremierGE Medical Systems, L.L.C. · Coil, magnetic resonance, specialty3 recalls0 reports
- MAGNETOM Skyra 10849580Siemens Healthcare GmbH · Coil, magnetic resonance, specialty2 recalls2 reports
- MAGNETOM Spectra 10837643Siemens Shenzhen Magnetic Resonance Ltd. · Coil, magnetic resonance, specialty1 recall1 report
- MR Coils dS Breast 16ch 3.0TPhilips Medical Systems Nederland B.V. · Coil, Magnetic Resonance, Specialty1 recall0 reports
- Invivo 45353028072Philips Ds North America LLC · Coil, magnetic resonance, specialty1 recall0 reports
- Philips dS Breast 7ch 3.0TPhilips Ds North America LLC · Coil, magnetic resonance, specialty1 recall0 reports
- NA 1.5T 4CH SMALL FLEX COILGe Healthcare(Tianjin) Company Limited · Coil, magnetic resonance, specialty1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page