Skip to content

Reform Pedicle Screw System 39-PA-1050Recalls & reported problems

Precision Spine, Inc. · ø10.5 x 50 mm POLYAXIAL SCREW · Class 2 · Implant

All models of Reform Pedicle Screw System (11)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00815362028957), received between Apr 18, 2022 and Apr 18, 2022 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 18, 2022Break

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2022, TO ADDRESS A BROKEN REFORM PEDICLE SCREW, ATTRIBUTED TO NON-FUSION. THE ORIGINAL IMPLANTATION DATE WAS NOT PROVIDED. THE EXISTING CONSTRUCT WAS REMOVED AND UPON ATTEMPTED INSERTION OF A OF A REFORM 10.5 X 50MM POLYAXIAL PEDICLE SCREW (39-PA-1050) IN A NEWLY PREPARED PEDICLE, THE TIP OF THE REFORM POLYAXIAL DRIVER ((B)(4)) BROKE OFF IN THE HEAD OF THE SCREW. THE SCREW WAS THEN REMOVED AND SUCCESSFULLY REPLACED WITH ANOTHER SCREW USING A SECOND DRIVER IN THE SET. SUBSEQUENTLY, WHILE IMPLANTING ANOTHER REFORM 10.5 X 50MM POLYAXIAL PEDICLE SCREW (39-PA-1050) THE SCREW WAS 85% SEATED WHEN THE HEAD OF THE SCREW SHEARED OFF AT THE NECK BETWEEN THE SCREW HEAD AND THREADED SHAFT. THE THREADED SHAFT WAS LEFT IN THE PATIENT'S BONE AND THE PEDICLE WAS LEFT UN-FIXATED. THE CONSTRUCT WAS SUCCESSFULLY COMPLETED WITH NO HARM TO THE PATIENT. THERE WAS A DELAY OF ONE HOUR BECAUSE OF THE REPORTED MALFUNCTIONS.

    Patient problems: Foreign Body In Patient

    MAUDE report 14138139 ↗

Other Thoracolumbosacral Pedicle Screw System devices

All Thoracolumbosacral Pedicle Screw System devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page