FilmArray® 2.0 Instrument FLM2-ASY-0001Recalls & reported problems

Biofire Diagnostics, LLC · IVD Instrument (FilmArray® 2.0 Instrument). · Class 2

All models of FilmArray 2.0 Instrument (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00815381020031), received between Jun 10, 2024 and Feb 13, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “False Negative Result” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 7, 2024Class IIOpen, Classified

    Some automated in vitro diagnostic modules and instruments manufactured or serviced between May 2022 and May 2024 were calibrated with internal equipment that may lead modules and instruments to report higher melting temperatures, which may lead to erroneous results including false positive Salmonella, Giardia lamblia, false positive/negative Norovirus, and false negative MRSA.

    Z-3122-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • False Negative Result2
  • Failure to Seal1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 13, 2025False Negative Result; False Negative Result

    SUMMARY: OSPEDALE CA' FONCELLO TREVISO (TREVISO, ITALY) REPORTED A POTENTIAL FALSE NEGATIVE VERONA INTEGRON-ENCODED METALLO-BETA-LACTAMASE (VIM) RESULT ON THE BIOFIRE BLOOD CULTURE IDENTIFICATION (BCID2) PANEL AFTER TESTING A PATIENT SAMPLE. THE PATIENT WAS NOT IMPACTED DUE TO THE BIOFIRE BCID2 PANEL RESULT. THE INVESTIGATION CONCLUDED THAT THE MOSTLY LIKELY CAUSE FOR THE DISCREPANT VIM RESULT WAS FILMARRAY 2.0 INSTRUMENT ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21380591 ↗
  • MalfunctionFeb 13, 2025False Negative Result; False Negative Result

    SUMMARY: OSPEDALE CA' FONCELLO TREVISO (TREVISO, ITALY) REPORTED A POTENTIAL FALSE NEGATIVE PSEUDOMONAS AERUGINOSA RESULT ON THE BIOFIRE BLOOD CULTURE IDENTIFICATION (BCID2) PANEL AFTER TESTING A PATIENT SAMPLE. THE PATIENT WAS NOT IMPACTED DUE TO THE BIOFIRE BCID2 PANEL RESULT. THE INVESTIGATION CONCLUDED THAT THE MOSTLY LIKELY CAUSE FOR THE DISCREPANT P. AERUGINOSA RESULT WAS FILMARRAY 2.0 INSTRUMENT ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21380536 ↗
  • MalfunctionJun 10, 2024Failure to Seal; Failure to Seal

    SUMMARY: (B)(6) REPORTED A POTENTIAL FALSE NEGATIVE NEISSERIA MENINGITIDIS RESULT ON FILMARRAY MENINGITIS/ENCEPHALITIS (ME) PANEL AFTER TESTING A PATIENT'S CEREBROSPINAL FLUID (CSF). DUE TO THE FILMARRAY ME PANEL RESULT, THE PATIENT'S TREATMENT WAS DELAYED. THE INVESTIGATION CONCLUDED THAT THE MOST LIKELY CAUSE FOR THE DISCREPANT RESULT WAS INSTRUMENT ERROR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19498946 ↗

Other Instrumentation For Clinical Multiplex Test Systems devices

All Instrumentation For Clinical Multiplex Test Systems devices →