Skip to content

FilmArray® 2.0 Instrument FLM2-ASY-0001-WRecalls & reported problems

Biofire Diagnostics, LLC · Refurbished IVD Instrument (FilmArray® 2.0 Instrument). · Class 2

All models of FilmArray 2.0 Instrument (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Aug 7, 2024Class IIOpen, Classified

    Some automated in vitro diagnostic modules and instruments manufactured or serviced between May 2022 and May 2024 were calibrated with internal equipment that may lead modules and instruments to report higher melting temperatures, which may lead to erroneous results including false positive Salmonella, Giardia lamblia, false positive/negative Norovirus, and false negative MRSA.

    Z-3122-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Other Instrumentation For Clinical Multiplex Test Systems devices

All Instrumentation For Clinical Multiplex Test Systems devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page