FilmArray® Meningitis/Encephalitis (ME) Panel RFIT-ASY-0119Recalls & reported problems
Biofire Diagnostics, LLC · IVD reagent kit containing 6 tests. · Class 2
All models of FilmArray Meningitis/Encephalitis (ME) Panel (2)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 2
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00815381020130), received between Sep 9, 2018 and Feb 15, 2019 — classified as 2 deaths.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings2
- False Negative Result1
- False Positive Result1
Patient problems
- Death2
- Meningitis1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
DeathFeb 15, 2019False Negative Result; Incorrect, Inadequate or Imprecise Result or Readings; False Negative Result; Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER (THROUGH (B)(4)) REPORTED A DISCREPANT CRYPTOCOCCUS NEOFORMANS/GATTII (C. NEOFORMANS/GATTII) RESULT ON THE FILMARRAY MENINGITIS/ENCEPHALITIS (ME) PANEL AFTER TESTING A CEREBROSPINAL FLUID (CSF) SAMPLE FROM A (B)(6), IMMUNOCOMPROMISED, MALE PATIENT. ACCORDING TO THE CUSTOMER, THE DISCREPANT FILMARRAY ME PANEL RESULT LED TO INAPPROPRIATE PATIENT TREATMENT, SUBSEQUENT PATIENT DETERIORATION AND DEATH. UPON INVESTIGATION, IT WAS FOUND THAT NO MALFUNCTION OCCURRED AND THE FILMARRAY ME PANEL SYSTEM WAS WORKING WITHIN SPECIFICATIONS. THE INVESTIGATION CONCLUDED THAT THE MOST LIKELY CAUSES FOR THE DISCREPANT RESULT WERE: SENSITIVITY/SPECIFICITY DIFFERENCES BETWEEN ANTIGEN AND PCR METHODS, AND/OR SAMPLE HANDLING ERROR.
Patient problems: Death
MAUDE report 8345345 ↗DeathSep 9, 2018False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings; False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings
(B)(6) REPORTED A DISCREPANT CRYPTOCOCCUS RESULT ON THE FILMARRAY MENINGITIS/ ENCEPHALITIS (ME) PANEL USING A PATIENT CEREBROSPINAL FLUID (CSF) SAMPLE. PATIENT CARE WAS AFFECTED DUE TO THE FILMARRAY RESULTS. UPON INVESTIGATION, NO MALFUNCTION OCCURRED AND THE SYSTEM WAS WORKING WITHIN SPECIFICATIONS. THE INVESTIGATION CONCLUDED THAT THE MOST LIKELY CAUSE FOR THE DISCREPANT RESULT WAS DUE TO LOW LEVEL ORGANISM IN THE SAMPLE AND COMPARATOR SENSITIVITY/SPECIFICITY DIFFERENCES. ON (B)(6) 2018, BIOFIRE RECEIVED ADDITIONAL INFORMATION FROM THE CUSTOMER INDICATING THAT THE PATIENT HAD DIED. BIOFIRE HAS REQUESTED ADDITIONAL INFORMATION ABOUT THE CAUSE OF DEATH, BUT HAS NOT RECEIVED ADDITIONAL INFORMATION. BIOFIRE HAS NO INFORMATION TO SUGGEST THAT USE OF THE FILMARRAY ME PANEL CAUSED OR CONTRIBUTED TO THE PATIENT'S DEATH.
Patient problems: Death; Meningitis
MAUDE report 7858817 ↗
Other Meningitis/Encephalitis Pathogen Multiplex Nucleic Acid Detection System devices
- FilmArray® Meningitis/Encephalitis (ME) Panel RFIT-ASY-0118Biofire Diagnostics, LLC · Meningitis/Encephalitis Pathogen Multiplex Nucleic Acid Detection System6 deaths30 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page