MiniBunion™ MinBunion Capital Fragment Guide Kit - SterileRecalls & reported problems
Crossroads Extremity Systems · The MiniBunion™ Implant System is indicated for stabilization and reconstruction of small bones of the hand, feet, wrist, ankles, fingers and toes. It consist of an intramedullary implant and screws. · Class 1
All models of MiniBunion (7)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00815432029297), received between Dec 14, 2023 and Jan 11, 2024 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device-Device Incompatibility4
- Defective Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 11, 2024Defective Device; Device-Device Incompatibility; Defective Device; Device-Device Incompatibility
IT WAS REPORTED THAT DURING SURGERY ON AN UNKNOWN DATE, THE 1.6MM K-WIRES THAT ARE SUPPOSED TO FIT IN THE MINIBUNION CAPITAL FRAGMENT GUIDE WOULD NOT FIT IN THE HOLES. THE STAFF TRIED TO USE SMALLER WIRES, BUT THOSE WERE NOT RIGID ENOUGH TO BE USED IN THE CASE. THEY CONTINUED THE CASE, AND THE PATIENT OUTCOME WAS NOT AFFECTED. THIS REPORT INVOLVES ONE MINIBUNION CAPITAL FRAG CTRL GUIDE STER. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18490877 ↗MalfunctionJan 3, 2024Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT SURGEON WAS PERFORMING A MINI-BUNION PROCEDURE ON (B)(6) 2023, AT APPROXIMATELY 8AM. SALES REP WAS COVERING THE CASE. WHILE ATTEMPTING TO UTILIZE THE CAPITAL FRAGMENT CONTROL GUIDE (AKA: VIKING SHIP) DR WAS UNABLE TO GET THE SUPPLIED 1.6MM K-WIRES INTO ANY OF THE HOLES. SMALLER K-WIRES WERE UTILIZED THAT FIT INTO THE HOLES. PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. THERE WAS NO PATIENT OUTCOME. THIS REPORT IS FOR ONE (1) MINIBUNION CAPITAL FRAG CTRL GUIDE STER. THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18434180 ↗MalfunctionDec 14, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT DURING A PROCEDURE ON (B)(6) 2023, THE HOLES IN THE CAPITAL FRAGMENT CONTROL GUIDE WERE FOUND TO BE TOO SMALL FOR THE PINS PROVIDED. THIS SAME ISSUE WAS FOUND ON BOTH IPS IN QUESTION. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING A SMALLER WIRE TO PASS THROUGH THE CAPITAL FRAGMENT GUIDE. THIS REPORT IS FOR A MINIBUNION CAPITAL FRAGMENT GUIDE KIT ¿ STERILE. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18326516 ↗MalfunctionDec 14, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT DURING A PROCEDURE ON (B)(6) 2023, THE HOLES IN THE CAPITAL FRAGMENT CONTROL GUIDE WERE FOUND TO BE TOO SMALL FOR THE PINS PROVIDED. THIS SAME ISSUE WAS FOUND ON BOTH IPS IN QUESTION. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING A SMALLER WIRE TO PASS THROUGH THE CAPITAL FRAGMENT GUIDE. THIS REPORT IS FOR A MINIBUNION CAPITAL FRAGMENT GUIDE KIT ¿ STERILE. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18326509 ↗
Other Guide, Surgical, Instrument devices
- NA 03.037.017Synthes GmbH · Guide, surgical, instrument346 reports
- EVOS 71174934Smith & Nephew, Inc. · Guide, surgical, instrument149 reports
- LCP 323.042Synthes GmbH · Guide, surgical, instrument129 reports
- DJO SURGICAL 342-10-705Encore Medical, L.P. · Guide, Surgical, Instrument1 recall51 reports
- DJO SURGICAL 341-10-707Encore Medical, L.P. · Guide, Surgical, Instrument81 reports
- NA 03.010.404Synthes GmbH · Guide, surgical, instrument73 reports
- DJO SURGICAL 341-14-708Encore Medical, L.P. · Guide, Surgical, Instrument1 recall22 reports
- DJO SURGICAL 341-10-711Encore Medical, L.P. · Guide, Surgical, Instrument1 recall16 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page