Skip to content

Penumbra Coil 400 400PAC35Recalls & reported problems

Penumbra, Inc. · PAC400, 35 cm · Class 2 · Implant

All models of Penumbra Coil 400 (120)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00815948021327), received between Jun 1, 2021 and Jun 1, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Premature Separation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Premature Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 1, 2021Premature Separation; Premature Separation

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE CAROTID-CAVERNOUS FISTULA USING A PAC400S, PENUMBRA COIL 400 (PC400S), AND PX SLIM DELIVERY MICROCATHETER (PX SLIM). DURING THE PROCEDURE, THE PHYSICIAN PLACED PC400S AND PAC400S IN THE TARGET LOCATION USING THE PX SLIM. NEXT, AFTER ADVANCING APPROXIMATELY FIFTY PERCENT OF A PAC400 INTO THE TARGET LOCATION AS THE FINAL COIL, THE PX SLIM BEGAN TO KICK BACK DUE TO THE LACK OF SPACE WITHIN THE FISTULA. THE PHYSICIAN MADE MULTIPLE ATTEMPTS TO RECAPTURE A SMALL SEGMENT OF THE PAC400, REPOSITIONED THE MICROCATHETER, AND ATTEMPTED TO REDEPLOY THE PAC400. HOWEVER, ON THE FINAL ATTEMPT TO REPOSITION THE PX SLIM, THE PHYSICIAN PULLED BACK ON THE PUSHER ASSEMBLY OF THE PAC400 AND NOTICED THE COIL HAD UNINTENTIONALLY DETACHED INSIDE THE PX SLIM. SUBSEQUENTLY, THE PHYSICIAN USED THE PUSHER ASSEMBLY OF THE PAC400 TO PUSH THE REMAINING COIL INTO PLACE. THE PHYSICIAN WAS SATISFIED WITH THE RESULTS AND THE PROCEDURE HAD ENDED AT THIS POINT. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11917983 ↗

Other DEVICE, NEUROVASCULAR EMBOLIZATION devices

All DEVICE, NEUROVASCULAR EMBOLIZATION devices →

Sources and updates