LP Coil System RBYPCLP03-BRecalls & reported problems
Penumbra, Inc. · Packing Coil LP, 3cm · Class 2 · Implant
All models of LP Coil System (38)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00815948026148), received between Jun 12, 2026 and Jun 12, 2026 — classified as 1 malfunction.
The most frequently reported specific device problem is “Premature Separation” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Premature Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 12, 2026Premature Separation
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE SPLENIC ARTERY USING PACKING COIL LPS, A LP SYSTEM DETACHMENT HANDLE (HANDLE), AND A NON-PENUMBRA TAPERED MICROCATHETER. IT WAS NOTED THAT THE PATIENT¿S VESSEL WAS TORTUOUS. A PACKING COIL LP WAS ADVANCED THROUGH THE MICROCATHETER INTO THE TARGET VESSEL; HOWEVER, THE PACKING COIL LP FAILED TO DETACH USING THE HANDLE. THE PHYSICIAN DECIDED TO REMOVE AND RESHEATH THE PACKING COIL LP. WHILE ATTEMPTING TO REMOVE THE PACKING COIL LP, THE PHYSICIAN EXPERIENCED RESISTANCE, AND THE EMBOLIZATION COIL UNINTENTIONALLY DETACHED WITHIN THE MICROCATHETER. THE PACKING COIL LP PUSHER ASSEMBLY WAS THEN USED TO PUSH THE EMBOLIZATION COIL INTO THE TARGET LOCATION. THE SAME ISSUE OCCURRED WITH TWO ADDITIONAL PACKING COIL LPS. BOTH EMBOLIZATION COILS WERE ALSO PUSHED INTO THE TARGET LOCATION USING THEIR PUSHER ASSEMBLIES. THE PROCEDURE WAS COMPLETED USING SEVEN NON-PENUMBRA DETACHABLE COILS AND THE SAME MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25486626 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page