Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00815948026148), received between Jun 12, 2026 and Jun 12, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Premature Separation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Premature Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 12, 2026Premature Separation

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE SPLENIC ARTERY USING PACKING COIL LPS, A LP SYSTEM DETACHMENT HANDLE (HANDLE), AND A NON-PENUMBRA TAPERED MICROCATHETER. IT WAS NOTED THAT THE PATIENT¿S VESSEL WAS TORTUOUS. A PACKING COIL LP WAS ADVANCED THROUGH THE MICROCATHETER INTO THE TARGET VESSEL; HOWEVER, THE PACKING COIL LP FAILED TO DETACH USING THE HANDLE. THE PHYSICIAN DECIDED TO REMOVE AND RESHEATH THE PACKING COIL LP. WHILE ATTEMPTING TO REMOVE THE PACKING COIL LP, THE PHYSICIAN EXPERIENCED RESISTANCE, AND THE EMBOLIZATION COIL UNINTENTIONALLY DETACHED WITHIN THE MICROCATHETER. THE PACKING COIL LP PUSHER ASSEMBLY WAS THEN USED TO PUSH THE EMBOLIZATION COIL INTO THE TARGET LOCATION. THE SAME ISSUE OCCURRED WITH TWO ADDITIONAL PACKING COIL LPS. BOTH EMBOLIZATION COILS WERE ALSO PUSHED INTO THE TARGET LOCATION USING THEIR PUSHER ASSEMBLIES. THE PROCEDURE WAS COMPLETED USING SEVEN NON-PENUMBRA DETACHABLE COILS AND THE SAME MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions