Penumbra, Inc. · Packing Coil LP, 15cm · Class 2 · Implant

All models of LP Coil System (38)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00815948026179), received between Nov 23, 2024 and Dec 11, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Premature Separation” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Premature Separation3
  • Material Deformation1
  • Migration1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Foreign Body In Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 11, 2025Premature Separation; Premature Separation

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE FOR AN ARTERIOVENOUS MALFORMATION (AVM) USING PACKING COIL LPS AND A NON-PENUMBRA MICROCATHETER. DURING THE PROCEDURE, THE PHYSICIAN PLACED SEVERAL PACKING COIL LPS IN THE TARGET LOCATION. AN ADDITIONAL PACKING COIL LP WAS ADVANCED TO THE TARGET LOCATION. THE PHYSICIAN DECIDED TO REPOSITION THE PACKING COIL LP IN THE TARGET LOCATION. WHILE RETRACTING THE PACKING COIL LP, THE EMBOLIZATION COIL UNINTENTIONALLY DETACHED IN THE MICROCATHETER. THE MICROCATHETER CONTAINING THE EMBOLIZATION COIL WAS REMOVED FROM THE PATIENT, AND THE PACKING COIL LP WAS FLUSHED OUT OF THE MICROCATHETER. THE PROCEDURE WAS COMPLETED WITH MORE PACKING COIL LPS AND THE SAME MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23776784 ↗
  • MalfunctionNov 3, 2025Premature Separation; Premature Separation

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE USING A PACKING COIL LP. DURING THE PROCEDURE, IT WAS REPORTED THAT THE COIL HAD UNINTENTIONALLY DETACHED. THE PHYSICIAN USED A SNARE DEVICE TO SUCCESSFULLY REMOVE THE PACKING COIL LP. THE PROCEDURE WAS COMPLETED USING A NEW PACKING COIL LP. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23458359 ↗
  • MalfunctionDec 13, 2024Material Deformation; Premature Separation; Material Deformation; Premature Separation

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE MIDDLE MENINGEAL ARTERY (MMA) USING PACKING COIL LPS AND A NON-PENUMBRA MICROCATHETER. DURING THE PROCEDURE, THE PHYSICIAN SUCCESSFULLY IMPLANTED ONE PACKING COIL LP INTO THE TARGET LOCATION. THE PHYSICIAN THEN ADVANCED ANOTHER PACKING COIL LP INTO THE TARGET LOCATION. WHILE RETRACTING THE PACKING COIL LP FOR REPOSITIONING, THE PHYSICIAN KINKED THE PUSHER ASSEMBLY OF THE PACKING COIL LP AND THE COIL UNINTENTIONALLY DETACHED. THE PHYSICIAN THEN USED THE PUSHER ASSEMBLY OF THE PACKING COIL LP TO SUCCESSFULLY PUSH THE REMAINING FIVE CENTIMETERS OF THE COIL OUT OF MICROCATHETER, AND INTO THE TARGET LOCATION. THE PROCEDURE WAS SUCCESSFULLY COMPLETED AT THIS POINT AND NO ADDITIONAL COILS WERE NECESSARY. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20926781 ↗
  • MalfunctionNov 23, 2024Migration; Migration

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE GASTRODUODENAL ARTERY (GDA) USING A PACKING COIL LP, RUBY COIL LPS, AND A NON-PENUMBRA MICROCATHETER. DURING THE PROCEDURE, THE PHYSICIAN PLACED THREE RUBY COIL LPS IN THE TARGET LOCATION USING THE MICROCATHETER. THE PHYSICIAN THEN SUCCESSFULLY PLACED ANOTHER PACKING COIL LP IN THE VESSEL. WHILE POWER INJECTING CONTRAST, THE LAST PACKING COIL LP PLACED WAS PUSHED FROM THE GDA TO THE RIGHT HEPATIC ARTERY. IT WAS REPORTED THAT THE PHYSICIAN DID NOT HAVE A SMALL ENOUGH SNARE DEVICE TO BE USED TO REMOVE THE PACKING COIL LP DUE TO THE TORTUOSITY AND DIAMETER OF THE VESSEL. THEREFORE, THE PACKING COIL LP WAS LEFT IN THE RIGHT HEPATIC ARTERY. THE PHYSICIAN DECIDED TO END THE PROCEDURE AT THIS POINT. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 20766097 ↗