Skip to content

MIDWAY 62MWAY115-ARecalls & reported problems

Penumbra, Inc. · MIDWAY 62 Delivery Catheter, 115cm, no CE · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00815948028197), received between Jun 25, 2024 and Jun 25, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Leak/Splash” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Leak/Splash1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 25, 2024Leak/Splash; Material Deformation; Leak/Splash; Material Deformation

    THE PATIENT WAS UNDERGOING A MEDICAL PROCEDURE IN THE INTERNAL CAROTID ARTERY (ICA) USING A MIDWAY 62 DELIVERY CATHETER. DURING THE PROCEDURE, THE PHYSICIAN ADVANCED THE MIDWAY 62 DELIVERY CATHETER TO THE TARGET LOCATION TO PERFORM A CONTRAST RUN. WHILE ATTEMPTING TO INJECT CONTRAST THROUGH THE MIDWAY 62 DELIVERY CATHETER, THE PHYSICIAN NOTICED THAT THE MIDWAY 62 DELIVERY CATHETER WAS LEAKING CONTRAST FROM A KINK ON THE STRAIN RELIEF. THEREFORE, THE MIDWAY 62 DELIVERY CATHETER WAS REMOVED. THE PROCEDURE WAS COMPLETED USING A NON-PENUMBRA ACCESS CATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19608734 ↗

Other Catheter, Percutaneous, Neurovasculature devices

All Catheter, Percutaneous, Neurovasculature devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page