Skip to content

Balloon Guide Catheter BOB-895-USRecalls & reported problems

Microvention, Inc. · BOBBY · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
11
Deaths reported
0
Injuries reported
4
Recalls naming it
1

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00842429108543), received between Oct 21, 2022 and Sep 19, 2025 — classified as 4 injuries, 7 malfunctions.

The most frequently reported specific device problem is “Flaked” (6 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 3 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 30, 2026Class IIOpen, Classified

    Due to an increase rate of balloon pinhole defects.

    Z-2912-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Flaked6
  • Adverse Event Without Identified Device or Use Problem4
  • Device Difficult to Setup or Prepare3
  • Device Handling Problem3
  • Contamination /Decontamination Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Obstruction/Occlusion2
  • Vascular Dissection2
  • Vasoconstriction2
  • Embolism/Embolus1
  • Muscle Weakness/Atrophy1
  • Reduced Blood Flow1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 19, 2025Flaked; Device Difficult to Setup or Prepare; Device Handling Problem; Flaked; Device Difficult to Setup or Prepare; Device Handling Problem

    IT WAS REPORTED THAT THIS MORNING WHILE PREPPING THE BOBBY FOR A CAROTID STENT CASE. THE TECH FOUND THAT THE BALLOON HAD LITTLE FLECKS OF MATERIAL STICKING OUT FROM THE BALLOON. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23097133 ↗
  • MalfunctionSep 19, 2025Flaked; Device Difficult to Setup or Prepare; Device Handling Problem; Flaked; Device Difficult to Setup or Prepare; Device Handling Problem

    IT WAS REPORTED THAT THIS MORNING WHILE PREPPING THE BOBBY FOR A CAROTID STENT CASE. THE TECH FOUND THAT THE BALLOON HAD LITTLE FLECKS OF MATERIAL STICKING OUT FROM THE BALLOON. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23095983 ↗
  • MalfunctionSep 18, 2025Flaked; Device Difficult to Setup or Prepare; Device Handling Problem; Flaked; Device Difficult to Setup or Prepare; Device Handling Problem

    IT WAS REPORTED THAT THIS MORNING WHILE PREPPING THE BOBBY FOR A CAROTID STENT CASE. THE TECH FOUND THAT THE BALLOON HAD LITTLE FLECKS OF MATERIAL STICKING OUT FROM THE BALLOON. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23092187 ↗
  • InjuryApr 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ADDITIONAL INFORMATION WAS PROVIDED THAT IS TO BE ADDED TO PREVIOUS SUBMISSION. INFORMATION RECEIVED NOTES THE FOLLOWING: EVENT SUMMARY : DURING THE ORIGINAL PROCEDURE ON (B)(6) 2024, THERE WAS THROMBUS SEEN DISTAL IN TWO MCA M2 BRANCHES FOLLOWING SUCCESSFUL THROMBECTOMY OF THE MCA M1 OCCLUSION. THIS MAY HAVE REPRESENTED PRE-EXISTING THROMBUS OF FRAGMENTATION OF THE ORIGINAL MCA M1 THROMBUS. THIS DISTAL MCA M2 BRANCH THROMBUS WAS TREATED WITH TWO ADDITIONAL PASSES WITH A STENT-RETRIEVER DURING THE ORIGINAL THROMBECTOMY. EVENT NAME : DISTAL EMBOLI EVENT : NEUROLOGICAL SEVERITY : MILD SAE? : NO UADE / USADE? : NO INTERVENTION - SPECIFY : CONTINUED THROMBECTOMY PASSES. OUTCOME - AE : RESOLVED WITHOUT SEQUELAE DATE OF RESOLUTION : ON (B)(6) 2024 RELATED TO: INITIAL THROMBECTOMY CLARIFICATION RECEIVED NOTES THE SITE UPDATED THE DISTAL EMBOLI IS POSSIBLY RELATED TO SOFIA AND BOBBY BALLOON GUIDE CATHETER, SOFIA DEVICE REMAINS AS CONCOMITANT DEVICE.

    Patient problems: Embolism/Embolus; Muscle Weakness/Atrophy; Obstruction/Occlusion

    MAUDE report 19198633 ↗
  • InjuryNov 15, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ADDITIONAL INFORMATION RECEIVED, SEE H10.

    Patient problems: Obstruction/Occlusion; Thrombosis/Thrombus

    MAUDE report 18145949 ↗
  • MalfunctionAug 16, 2023Flaked; Flaked

    IT WAS REPORTED THAT THE BOBBY WAS INFLATED PRIOR TO USE AND WAS OBSERVED TO HAVE MATERIAL STICKING OUT FROM THE BALLOON. THE BALLOON WAS NOT USED IN THE PATIENT. NO PATIENT HARM OR INJURY OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17561270 ↗
  • MalfunctionAug 15, 2023Flaked; Flaked

    IT WAS REPORTED THAT THE BOBBY WAS INFLATED PRIOR TO USE AND WAS OBSERVED TO HAVE MATERIAL STICKING OUT FROM THE BALLOON. THE BALLOON WAS NOT USED IN THE PATIENT. NO PATIENT HARM OR INJURY OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17553458 ↗
  • InjuryApr 28, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED, THE PATIENT PRESENTED WITH A CERVICAL CAROTID OCCLUSION SECONDARY TO A SEVERE DISSECTION. REPORTEDLY, THE USE OF THE BOBBY BALLOON CATHETER WORSENED THE DISSECTION AND CAUSED VASOSPASM. THE PATIENT WAS DISCHARGED HOME.

    Patient problems: Vascular Dissection; Vasoconstriction

    MAUDE report 16838834 ↗
  • InjuryApr 28, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED, THE PATIENT PRESENTED WITH OCCLUSION IN THE LEFT M2 SEGMENT OF THE MCA. REPORTEDLY, THE USE OF THE BOBBY BALLOON GUIDE CATHETER CAUSED A SEVERE DISSECTION AND VASOSPASM WITH FLOW LIMITATION IN THE CERVICAL ICA. PATIENT UNDERWENT A HEMICRANIECTOMY AND IS NOW IN INPATIENT REHAB.

    Patient problems: Reduced Blood Flow; Vascular Dissection; Vasoconstriction

    MAUDE report 16838031 ↗
  • MalfunctionJan 5, 2023Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT DURING PREPARATION, THE PHYSICIAN NOTICED A FOREIGN MATERIAL WAS STICKING OUTSIDE OF THE BALLOON. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16107926 ↗
  • MalfunctionOct 21, 2022Flaked; Flaked

    IT WAS REPORTED THAT DURING PREPARATION, A BALLOON CATHETER'S HYDROPHILIC COATING WAS OBSERVED TO BE "COMING APART." THE BALLOON WAS SET ASIDE AND ANOTHER WAS USED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15655571 ↗