Penumbra, Inc. · SwiftPAC Coil, 20 cm · Class 2 · Implant

All models of Swift Coil System (11)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00815948029675), received between Jul 24, 2025 and Oct 30, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Premature Separation” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Premature Separation3
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 30, 2025Physical Resistance/Sticking; Premature Separation; Physical Resistance/Sticking; Premature Separation

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE MIDDLE MENINGEAL ARTERY (MMA) USING A SWIFTPAC COIL (SWIFTPAC), A BENCHMARK 6F 071 DELIVERY CATHETER (BENCHMARK), A MIDWAY 43 DELIVERY CATHETER (MIDWAY 43), A NON-PENUMBRA MICROCATHETER, AND A GUIDEWIRE. DURING THE PROCEDURE, THE PHYSICIAN EXPERIENCED RESISTANCE WITH THE SWIFTPAC. THE EMBOLIZATION COIL UNINTENTIONALLY DETACHED PARTIALLY IN THE TARGET VESSEL AND THE MICROCATHETER. THE MICROCATHETER CONTAINING THE DETACHED EMBOLIZATION COIL WAS REMOVED. THE PROCEDURE WAS COMPLETED USING A NEW SWIFTPAC AND THE SAME MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23426145 ↗
  • MalfunctionAug 27, 2025Premature Separation; Premature Separation

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE MIDDLE MENINGEAL ARTERY (MMA) USING A SWIFTPAC COIL (SWIFTPAC), A NON-PENUMBRA MICROCATHETER, AND A GUIDEWIRE. DURING THE PROCEDURE, WHILE RETRACTING THE SWIFTPAC TO REPOSITION IT, THE EMBOLIZATION COIL UNINTENTIONALLY DETACHED PARTIALLY IN THE MMA AND PARTIALLY IN THE INTERNAL CAROTID ARTERY (ICA). THE PHYSICIAN ATTEMPTED TO USE A NON-PENUMBRA STENT RETRIEVER TO REMOVE THE DETACHED EMBOLIZATION WITHOUT SUCCESS. A SECOND NON-PENUMBRA STENT RETRIEVER WAS THEN USED TO SUCCESSFULLY REMOVE THE DETACHED EMBOLIZATION COIL. THE PROCEDURE WAS COMPLETED USING EMBOLIZATION PARTICLES AND THE SAME MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22903397 ↗
  • MalfunctionJul 24, 2025Premature Separation; Premature Separation

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE MIDDLE MENINGEAL ARTERY (MMA) USING A SWIFTPAC COIL (SWIFTPAC) AND A NON-PENUMBRA MICROCATHETER. DURING THE PROCEDURE, WHILE ADVANCING THE SWIFTPAC THROUGH THE MICROCATHETER TO THE TARGET LOCATION, THE SWIFTPAC UNINTENTIONALLY DETACHED WITHIN THE MICROCATHETER. THE MICROCATHETER CONTAINING DETACHED SWIFTPAC WAS REMOVED. THE PROCEDURE WAS COMPLETED USING ANOTHER SWIFTPAC AND ANOTHER NON-PENUMBRA MICROCATHETER (HEADWAY DUO 156CM). THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22592946 ↗