Skip to content

ForaCare GD20Recalls & reported problems

Fora Care Inc. · Blood Glucose Monitoring System · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00816042010804), received between Nov 1, 2023 and Nov 1, 2023 — classified as 1 malfunction.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 1, 2023

    THE GLUCOMETER HAD NORMAL READINGS WHEN TESTING WITH THE CONTROL SOLUTION, HOWEVER, WHEN TESTING ON A PATIENT, THE RESULTS WERE VERY HIGH. OUR INITIAL READING WAS 589. THE HOSPITAL TESTED THE PATIENT BLOOD GLUCOSE WITH THEIR MACHINE AND GOT 113. WE DID A RETEST WITH THE BLOOD FROM THE SAME FINGER STICK ON OUR MACHINE AND GOT A SECONDARY READING OF 167. THE HOSPITAL DID A RETEST AND GOT 112.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18059500 ↗

Other System, Test, Blood Glucose, Over The Counter devices

All System, Test, Blood Glucose, Over The Counter devices →