NextSeq 550Dx Sequencing System 20005715Recalls & reported problems
Illumina, Inc. · The Illumina NextSeq 550Dx instrument is intended for sequencing DNA libraries prepared by adding sample indices and capture sequences to amplified targets. These libraries are captured on a flow cell and sequenced on the instrument using sequencing by synthesis (SBS) chemistry. SBS chemistry uses a reversible-terminator method to detect single nucleotide bases incorporated into growing DNA strands. Instrument software performs image analysis and base calling, assigning a quality score to each base for each sequencing cycle. When primary analysis finishes, secondary analysis processes base calls on the instrument. The NextSeq 550Dx instrument uses different secondary analysis modules depending on the workflow. For the germline or somatic variant calling modules, data processing includes de-multiplexing, FASTQ file generation, alignment, variant calling, and generation of variant call format (VCF) files containing information regarding variants found at specific positions in a reference genome. · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 4
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00816270020125), received between Nov 23, 2021 and Jun 6, 2023 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Defective Component” (3 reports).
4 FDA recall records name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Oct 28, 2025Open, Classified
An instrument intended for sequencing DNA libraries with in vitro diagnostic assays is used with reagent kits including flow cells that have flow cell gaskets that have a quality issue, which may lead to flow cell leaks during runs, which could cause run failures, that may result in: No/insufficient data output, instrument fluidics clogs, toxic substance exposure to user.
Z-0884-2026 · Root cause: Vendor change control · FDA record ↗
- Apr 5, 2023Completed
Cybersecurity vulnerability concerning the software used for sequencing instruments.
Z-1976-2023 · Root cause: Software Manufacturing/Software Deployment · FDA record ↗
- May 3, 2022Completed
cybersecurity vulnerability
Z-1260-2022 · Root cause: Software in the Use Environment · FDA record ↗
- Feb 22, 2022Completed
There is a potential that the DNA sequence analyzer may short circuit shortly after disconnection and reconnection of circuit board or imaging module due to an insufficient reconnection of the ribbon cable. The short circuit results in the device being inoperable.
Z-0854-2022 · Root cause: Device Design · FDA record ↗
Most reported problems
Device problems
- Defective Component3
- Leak/Splash1
Patient problems
- Chemical Exposure2
- Chest Pain1
- Sore Throat1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 6, 2023Leak/Splash; Defective Component; Leak/Splash; Defective Component
THE IMPACTED PRODUCT IS THE NEXTSEQ 550 SEQUENCING SYSTEM AND SERIAL NUMBER, WHICH IS A RESEARCH USE ONLY (RUO) PRODUCT. THIS PRODUCT IS SIMILAR TO THE NEXTSEQ¿ 550DX INSTRUMENT IN VITRO DIAGNOSTIC (IVD). THE SEQUENCING BY SYNTHESIS (SBS) CHEMISTRY FOR THE RUO INSTRUMET IS SIMILAR TO THE IVD INSTRUMENT, THEREFORE, THIS EVENT IS BEING REPORTED AS AN MDR. NOTE: INDIVIDUAL FORMULATIONS MAY DIFFER SOMEWHAT. THE NEXTSEQ¿ 550DX INSTRUMENT IS INTENDED FOR SEQUENCING DNA LIBRARIES WITH IN VITRO DIAGNOSTIC ASSAYS. FOR ITS INPUT, THE NEXTSEQ¿ 550DX USES LIBRARIES GENERATED FROM DNA WHERE SAMPLE INDEXES AND CAPTURE SEQUENCES ARE ADDED TO AMPLIFIED TARGETS. SAMPLE LIBRARIES ARE CAPTURED ON A FLOW CELL AND SEQUENCED ON THE INSTRUMENT USING SEQUENCING BY SYNTHESIS (SBS) CHEMISTRY. SBS CHEMISTRY USES A REVERSIBLE-TERMINATOR METHOD TO DETECT FLUORESCENTLY LABELED SINGLE NUCLEOTIDE BASES AS THEY ARE INCORPORATED INTO GROWING DNA STRANDS. ON (B)(6) 2023, A CUSTOMER REPORTED THAT THEY OBSERVED A LEAK COMING FROM THEIR NEXTSEQ 550. THE LEAK WAS DISCOVERED ABOUT 75% OF THE WAY THROUGH THE RUN. THE RUN PROCEEDED FOR ABOUT 10 HOURS BEFORE THE LEAK WAS DISCOVERED. IT APPEARS THE WASTE DISPOSAL SPOUT INSID
Patient problems: Chemical Exposure
MAUDE report 17075873 ↗MalfunctionJun 15, 2022Defective Component; Defective Component
THE IMPACTED PRODUCT IS THE NEXTSEQ 500 SEQUENCING SYSTEM AND SERIAL NUMBER (B)(4), WHICH IS A RESEARCH USE ONLY (RUO) PRODUCT. THIS PRODUCT IS SIMILAR TO THE NEXTSEQ 550DX INSTRUMENT IN VITRO DIAGNOSTIC (IVD). THE SEQUENCING BY SYNTHESIS (SBS) CHEMISTRY FOR THE RUO INSTRUMENT IS SIMILAR TO THE IVD INSTRUMENT, THEREFORE, THIS EVENT IS BEING REPORTED AS AN MDR. NOTE: INDIVIDUAL FORMULATIONS MAY DIFFER SOMEWHAT. THE NEXTSEQ¿ 550DX INSTRUMENT IS INTENDED FOR SEQUENCING DNA LIBRARIES WITH IN VITRO DIAGNOSTIC ASSAYS. FOR ITS INPUT, THE NEXTSEQ¿ 550DX USES LIBRARIES GENERATED FROM DNA WHERE SAMPLE INDEXES AND CAPTURE SEQUENCES ARE ADDED TO AMPLIFIED TARGETS. SAMPLE LIBRARIES ARE CAPTURED ON A FLOW CELL AND SEQUENCED ON THE INSTRUMENT USING SEQUENCING BY SYNTHESIS (SBS) CHEMISTRY. SBS CHEMISTRY USES A REVERSIBLE-TERMINATOR METHOD TO DETECT FLUORESCENTLY LABELED SINGLE NUCLEOTIDE BASES AS THEY ARE INCORPORATED INTO GROWING DNA STRANDS. ON (B)(6) 2022, A CUSTOMER REPORTED THAT AFTER THE RUN WAS COMPLETED ON THEIR NEXTSEQ 500, THE WASTE HAD LEAKED AS THE WASTE BOTTLE WAS NOT PLACED CORRECTLY DURING THE RUN. NO HARM OR INJURY WAS REPORTED, BUT THE CUSTOMER REPORTED BEING CONCERNED ABOUT SLIPP
Patient problems: Chemical Exposure
MAUDE report 14691958 ↗InjuryNov 23, 2021Defective Component; Defective Component
THE IMPACTED PRODUCT IS THE NEXTSEQ 550 SEQUENCING SYSTEM (15067451) AND SERIAL NUMBER (B)(4), WHICH IS A RESEARCH USE ONLY (RUO) PRODUCT. THIS PRODUCT IS SIMILAR TO THE NEXTSEQ¿ 550DX INSTRUMENT IN VITRO DIAGNOSTIC (IVD). THE NEXTSEQ¿ 550DX INSTRUMENT IS INTENDED FOR SEQUENCING DNA LIBRARIES WITH IN VITRO DIAGNOSTIC ASSAYS. FOR ITS INPUT, THE NEXTSEQ¿ 550DX USES LIBRARIES GENERATED FROM DNA WHERE SAMPLE INDEXES AND CAPTURE SEQUENCES ARE ADDED TO AMPLIFIED TARGETS. SAMPLE LIBRARIES ARE CAPTURED ON A FLOW CELL AND SEQUENCED ON THE INSTRUMENT USING SEQUENCING BY SYNTHESIS (SBS) CHEMISTRY. SBS CHEMISTRY USES A REVERSIBLE-TERMINATOR METHOD TO DETECT FLUORESCENTLY LABELED SINGLE NUCLEOTIDE BASES AS THEY ARE INCORPORATED INTO GROWING DNA STRANDS. THE SEQUENCING BY SYNTHESIS (SBS) CHEMISTRY FOR THE RUO INSTRUMENT IS SIMILAR TO THE IVD INSTRUMENT, THEREFORE, THIS EVENT IS BEING REPORTED AS AN MDR. NOTE: INDIVIDUAL FORMULATIONS MAY DIFFER SOMEWHAT. ON (B)(6) 2021, A CUSTOMER, (B)(6), REPORTED THAT AFTER COMPLETION OF THE SEQUENCING RUN ON ILLUMINA NEXTSEQ 550 INSTRUMENT (RUO), A PURPLE SOLUTION WAS FOUND ON THE FLOOR. THE LABORATORY WAS EVACUATED. THE SAFETY TEAM WAS ON SITE AND CLEANED THE
Patient problems: Chest Pain; Sore Throat
MAUDE report 12865857 ↗
Other High Throughput Dna Sequence Analyzer devices
- MiSeqDx Instrument 15036706Illumina, Inc. · High Throughput Dna Sequence Analyzer2 recalls1 report
- NovaSeq™ 6000Dx 20068232Illumina, Inc. · High Throughput Dna Sequence Analyzer1 recall0 reports
- Torrent Suite DX SW, Installer V5.14 A52422Life Technologies Holdings Pte. Ltd. · High throughput DNA sequence analyzer1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.