Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00816349011832), received between Jun 30, 2023 and Mar 25, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Material Split, Cut or Torn” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Split, Cut or Torn2
- Device Damaged by Another Device1
- Material Deformation1
Patient problems
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 25, 2025Device Damaged by Another Device; Material Deformation; Material Split, Cut or Torn; Device Damaged by Another Device; Material Deformation; Material Split, Cut or Torn
EVENT DESCRIPTION: DURING PCNL COATING CAME OFF WIRE AND IN PATIENT. MD WAS ABLE TO RETRIEVE RETAINED WIRE. WHAT TROUBLESHOOTING STEPS TOOK PLACE? WHAT TROUBLESHOOTING STEPS, IF ANY, RESOLVED THE ISSUE? N/A. WHAT IS THE NEXT COURSE OF ACTION? WIRE REPLACED. PATIENT PRESENT AT TIME OF EVENT? YES. PATIENT COMPLICATIONS: NO PATIENT COMPLICATIONS. TIME OF THE EVENT: DURING PROCEDURE. WHAT WAS THE OUTCOME OF THE PROCEDURE? PROCEDURE COMPLETED USING AN ALTERNATE DEVICE. IMAGES RECEIVED 17 MARCH 2025. ADDITIONAL INFORMATION PROVIDED 24 MARCH 2025: 1. LISTING AND MODEL OF OTHER DEVICES USED DURING THE PROCEDURE, INCLUDE THE MODEL, BRAND NAME, SIZES, ETC. NEPHROSCOPE, UNSURE OF MANUFACTURER. 2. WAS THE ZIPWIRE ADVANCED THROUGH A METAL CANNULA OR NEEDLE? NEPHROSCOPE SHEATH. 3. DID THIS EVENT OCCUR DURING INSERTION, ADVANCEMENT, OR WITHDRAWAL? ONGOING USE. 4. DID THE PHYSICIAN ENCOUNTER ANY RESISTANCE WHILE INSERTING, ADVANCING, OR REMOVING THE ZIPWIRE? UNKNOWN. 5. WHAT METHOD WAS USED TO RETRIEVE THE "RETAINED WIRE"? UNKNOWN. 6. ARE THERE EFFORTS TO OBTAIN THE LOT NUMBER FOR THE COMPLAINT DEVICE? YES, SENT PACKAGING BACK WITH RETURN. 7. WHAT IS THE ANATOMY LOCATION OF THE DEVICE? KIDNEY/URET
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21688077 ↗MalfunctionJun 30, 2023Material Split, Cut or Torn; Material Split, Cut or Torn
PER MEDWATCH, IT WAS REPORTED THAT: SAFETY WIRE WAS REMOVED. INSPECTION OF GLIDEWIRE AT THE END OF THE CASE SHOWED 2 MILLIMETER(S) (MM) TIP OF GLIDEWIRE BROKE OFF. NOTE FROM DISTRIBUTOR RECEIVED (B)(6), 2023: WE WILL NOT BE ABLE TO SEND A FOLLOW UP ON THIS COMPLAINT SINCE INITIAL REPORTER AND HEALTH CARE FACILITY ARE UNKNOWN. THE DEVICE IS RETURNING; HOWEVER, IT HAS NOT BEEN RECEIVED.
Patient problems: Insufficient Information
MAUDE report 17239266 ↗
Other Stylet, Ureteral devices
- ZIPwire™ M0066802220Lake Region Medical · Stylet, Ureteral46 reports
- ZIPwire™ M006630222B0Lake Region Medical · Stylet, Ureteral21 reports
- OLYMPUS GWH3505RGyrus ACMI, LLC · STYLET, URETERAL16 reports
- ZIPwire™ M006630205B0Lake Region Medical · Stylet, Ureteral10 reports
- ZIPwire™ M0066802050Lake Region Medical · Stylet, Ureteral9 reports
- ZIPwire™ M0066802080Lake Region Medical · Stylet, Ureteral9 reports
- ZIPwire™ M006630209B0Lake Region Medical · Stylet, Ureteral5 reports
- OLYMPUS GWH3505RAGyrus ACMI, LLC · STYLET, URETERAL5 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.