ZIPwire™ M0066802160Recalls & reported problems
Lake Region Medical · Urological guidewire · Class 1
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00816349011948), received between Nov 17, 2022 and Jan 8, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Device Damaged by Another Device” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Device Damaged by Another Device1
- Insufficient Device Problem Information1
- Material Split, Cut or Torn1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 8, 2025Device Damaged by Another Device; Material Split, Cut or Torn; Device Damaged by Another Device; Material Split, Cut or Torn
EVENT DESCRIPTION: ECN EVENT DESCRIPTION: FROM CUSTOMER, " WE USE THE ZIP WIRES IN THE VASCULAR SURGERY PRACTICE AND THE ONE WE USED TODAY WAS DEFECTIVE. THE SHEATH DETACHED FROM ITSELF AND SLID DOWN THE WIRE, WHICH WOULD HAVE BEEN A HUGE CONCERN IF IT DETACHED COMPLETELY IN THE PATIENTS VEIN." WHAT TROUBLESHOOTING STEPS TOOK PLACE? WHAT TROUBLESHOOTING STEPS, IF ANY, RESOLVED THE ISSUE?: UNAVAILABLE WHAT IS THE NEXT COURSE OF ACTION?: UKN PATIENT PRESENT AT TIME OF EVENT?: YES PATIENT COMPLICATIONS: NO PATIENT COMPLICATIONS TIME OF THE EVENT: DURING PROCEDURE. ADDITIONAL INFORMATION PROVIDED 8 JANUARY 2025: 1. WHAT WAS THE NAME OF THE PROCEDURE(S) PERFORMED? RADIOFREQUENCY ABLATION. 2. IS THERE A PHOTO OF THE COMPLAINT DEVICE? A PICTURE HAS BEEN ATTACHED TO THIS EMAIL. 3. WHAT WAS THE OUTCOME OF THE PROCEDURE (COMPLETED WITH THIS DEVICE, COMPLETED WITH ANOTHER OF THE SAME DEVICE, COMPLETED WITH A DIFFERENT DEVICE, ETC.)? COMPLETED WITH THIS DEVICE; DEFECT WAS NOTICED UPON REMOVING THE DEVICE FROM THE PATIENT. 4. WHAT WAS THE PATIENT OUTCOME FOLLOWING THE PROCEDURE (GOOD, STABLE ETC.)? GOOD. 5. WHAT WAS THE ANATOMY LOCATION OF THE DEVICE? LOWER EXTREMITY 6. WAS THE ZIPWIRE ADVANCED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21106991 ↗MalfunctionNov 17, 2022Insufficient Device Problem Information; Insufficient Device Problem Information
EVENT DESCRIPTION: PER EMAIL, IT WAS REPORTED THAT: WE RECEIVED A PRODUCT FAILURE RECENTLY OF A GUIDEWIRE BREAKING DURING A PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15816445 ↗
Other Stylet, Ureteral devices
- ZIPwire™ M0066802220Lake Region Medical · Stylet, Ureteral46 reports
- ZIPwire™ M006630222B0Lake Region Medical · Stylet, Ureteral21 reports
- OLYMPUS GWH3505RGyrus ACMI, LLC · STYLET, URETERAL16 reports
- ZIPwire™ M006630205B0Lake Region Medical · Stylet, Ureteral10 reports
- ZIPwire™ M0066802050Lake Region Medical · Stylet, Ureteral9 reports
- ZIPwire™ M0066802080Lake Region Medical · Stylet, Ureteral9 reports
- ZIPwire™ M006630209B0Lake Region Medical · Stylet, Ureteral5 reports
- OLYMPUS GWH3505RAGyrus ACMI, LLC · STYLET, URETERAL5 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.