Circulo DAIBW-001Recalls & reported problems
Lake Region Medical · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00816349012662), received between Dec 13, 2022 and Dec 3, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Entrapment of Device” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Entrapment of Device2
- Material Split, Cut or Torn2
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 3, 2025Entrapment of Device; Entrapment of Device
ATTEMPT TO CROSS THE VALVE WITH THE DELIVERY SYSTEM AND WAS UNSUCCESSFUL AS THE SYSTEM WAS HITTING CALCIUM. ATTEMPTED TO REMOVE THE DELIVERY SYSTEM TO CHANGE OUT THE WIRE AND THE CIRCULO XS WIRE WAS STUCK IN THE DELIVERY SYSTEM. THE PHYSICIAN HAD TO PULL THE ENTIRE SYSTEM OUT WITH THE WIRE AND RE-CROSS THE VALVE WITH A STRAIGHT WIRE. 2ND ATTEMPT MADE TO CROSS THE VALVE WITH THE DELIVERY SYSTEM AND UNSUCCESSFUL AGAIN. THE PHYSICIAN ABORTED THE CASE WITH NAVITOR AND HAD EDWARDS COMPLETE THE CASE WITH A SAPIEN IMPLANT, USING THE STEERABLE SHEATH TO CROSS THE VALVE. ADDITIONAL INFORMATION: QUESTION: HAS THE HEALTHCARE PROFESSIONAL STATED THAT THEY BELIEVE THE PATIENT OR USER EXPERIENCED ADVERSE HEALTH CONSEQUENCES DUE TO THE PERFORMANCE OF THE PRODUCT? ANSWER: NO. QUESTION: PATIENT STATUS? ANSWER: STABLE AND DISCHARGED QUESTION. WERE ALL STEPS PERFORMED PER IFU? ANSWER: YES QUESTION: WAS PRE-BAV PERFORMED? ANSWER: YES 24 TRUE DILATION QUESTION: WAS THERE ANY DAMAGE NOTED TO THE CIRCULO WIRE PRIOR TO OR AFTER USE? ANSWER: YES - STUCK IN THE DELIVERY SYSTEM AND SENT BACK QUESTION: WAS THERE DAMAGE TO THE FIRST FLEXNAV DELIVERY SYSTEM? ANSWER: NO, OTHER THAN THE WIRE STUCK INSIDE QUESTION
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23706455 ↗MalfunctionJul 10, 2025Entrapment of Device; Entrapment of Device
ADDITIONAL INFORMATION: QUESTION: CAN YOU PLEASE CONFIRM WAS A CATHETER ADVANCED INTO THE TREATMENT AREA OVER THE CIRCULOTM GUIDEWIRE PRIOR TO WITHDRAWING THE WIRE. ANSWER: THE CATHETER WAS ADVANCED OVER THE CIRCULO WIRE INTO THE TREATMENT AREA. THE WIRE WAS MANIPULATED WITHOUT ISSUE IN THE TREATMENT AREA. THE CATHETER WAS REMOVED FROM TREATMENT AREA WITHOUT ISSUE WITH WIRE STILL IN THE HEART. ENTIRE SYSTEM WAS REMOVED TOGETHER FROM BODY WITHOUT ISSUE. ON THE TABLE, AFTER VALVE WAS DEPLOYED, THE WIRE WAS UNABLE TO BE PULLED THROUGH FLEXNAV SYSTEM AND GOT STUCK. WAS NOT ABLE TO BE PUSHED THROUGH THE SYSTEM EITHER AT THIS POINT. QUESTION: CAN YOU PLEASE CONFIRM WAS THE CIRCULOTM GUIDEWIRE PULLED BACK COMPLETELY INTO THE CATHETER? ANSWER: IN THE BODY NO IT WAS NOT. ON THE TABLE IT WAS PULLED BACK INTO THE CATHETER AND THIS IS WHEN IT GOT STUCK. QUESTION: CAN YOU PLEASE CONFIRM WAS THE CIRCULOTM GUIDEWIRE WITHDRAWN FROM THE CATHETER. ANSWER: THE CATHETER WAS REMOVED FROM TREATMENT AREA WITHOUT ISSUE WITH WIRE STILL IN THE HEART. ENTIRE SYSTEM WAS REMOVED TOGETHER FROM BODY WITHOUT ISSUE. ON THE TABLE, AFTER VALVE WAS DEPLOYED, THE WIRE WAS UNABLE TO BE PULLED THROUGH FLEXNAV SYSTEM AND
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22468558 ↗MalfunctionAug 26, 2024Material Split, Cut or Torn; Material Split, Cut or Torn
INFORMATION PROVIDED 30AUG2024: QUESTION: HOW THIS PROCEDURE WAS COMPLETED (I.E., ANOTHER CIRCULO WIRE, A DIFFERENT MODEL GUIDEWIRE, OR WAS THE ORIGINAL CIRCULO WIRE USED TO COMPLETE THE PROCEDURE)? ANSWER: WIRE WAS REMOVED THROUGH SGC DILATOR. SGC WAS ALREADY ACROSS SEPTUM SO WIRE WAS MERELY REMOVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20062952 ↗MalfunctionDec 13, 2022Material Split, Cut or Torn; Material Split, Cut or Torn
USED CIRCULO XS WIRE TO DO PREBAV AND DEPLOY 27MM PORTICO VALVE. UPON ATTEMPT TO REMOVE CIRCULO WIRE FOR A WIRE EXCHANGE, THE WIRE BECAME STUCK IN THE DELIVERY SYSTEM. UPON DIFFICULT REMOVAL, WIRE CAME OUT FRAYED WITH THE TIP STILL ATTACHED. ADDITIONAL INFORMATION RECEIVED NOVEMBER 22, 2022: Q1. WAS THERE ANY DAMAGE NOTED TO THE CIRCULO GUIDEWIRE PRIOR TO USE? A1. NO. Q2. WAS THERE ANY RESISTANCE NOTED DURING THE INITIAL PLACEMENT OF THE CIRCULO GUIDEWIRE WITHIN THE PATIENT OR AT ANY TIME PRIOR TO THE RESISTANCE NOTED WITH THE DELIVERY SYSTEM AND THE GUIDEWIRE? A2. NO. Q3. WHAT WAS THE MODEL OF THE BALLOON CATHETER USED IN THE PROCEDURE TO PERFORM THE BAV? A3. TRUE BALLOON. Q4. WAS BAV SUCCESSFULLY COMPLETED OVER CIRCULO WIRE? A4. YES. Q5. WAS BAV CATHETER REMOVED SUCCESSFULLY OVER CIRCULO WIRE? A5. YES. Q6. IF THE BAV CATHETER WAS REMOVED OVER THE CIRCULO WIRE, WAS THERE ANY RESISTANCE FELT AT ANY POINT? A6. NO. Q7. WHAT WAS THE REASON FOR THE WIRE EXCHANGE? A7. POST VALVE PLACEMENT, DIFFERENT WIRE NEED TO MAINTAIN VESSEL ACCESS FOR VESSEL CLOSURE DEVICES. Q8. THE EVENT DESCRIPTION MENTIONS THE PORTICO VALVE. PLEASE CONFIRM IF THE ABBOTT FLEX NAV DELIVERY SYSTEM WAS USED OR A DIFF
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15973206 ↗
Other Wire, Guide, Catheter devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page