Skip to content

N/A FC1SPRecalls & reported problems

Stryker Corporation · Single Fiber Optic Cable 10ft, Pink · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00816728020202), received between Aug 6, 2018 and May 27, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Overheating of Device” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Overheating of Device2
  • Temperature Problem2
  • Use of Device Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient2
  • Burn(s)1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 27, 2025Overheating of Device; Temperature Problem; Overheating of Device; Temperature Problem

    IT WAS REPORTED THAT THE LIGHT CORD WAS PLUGGED INTO A STRYKER L11 LIGHT SOURCE AND BECAME VERY HOT AFTER ONLY 30 MINUTES OF BEING PLUGGED IN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22091106 ↗
  • MalfunctionJan 3, 2025Temperature Problem; Temperature Problem

    IT WAS REPORTED THAT THERE WAS A THERMAL EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21072928 ↗
  • MalfunctionJun 11, 2021Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE AT THE USER FACILITY, WHILE USING THE DEVICE THE PATIENT EXPERIENCED SECOND DEGREE BURNS. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: Burn(s)

    MAUDE report 11983124 ↗
  • MalfunctionAug 15, 2018Use of Device Problem; Use of Device Problem

    IT WAS REPORTED DURING A HIP REPLACEMENT PROCEDURE, THE FIBER OPTIC CABLE MAY HAVE BEEN PLACED ON THE DRAPE OF THE PATIENT. DURING USE, THE TEMPERATURE OF THE FIBER OPTIC CORD FELT HOT AND LEFT A BURN MARK ON THE DRAPE. THE CASE WAS COMPLETED WITHOUT INCIDENT OR PRODUCT MALFUNCTION. THERE WAS NO PATIENT INJURY INVOLVED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7784680 ↗
  • MalfunctionAug 6, 2018Use of Device Problem; Use of Device Problem

    DURING A HIP REPLACEMENT PROCEDURE, IT WAS OBSERVED THAT THE DRAPE HAD A BURN MARK WHERE THE FIBER OPTIC CABLE HAD BEEN RESTING UPON IT. THERE WAS NO INJURY TO THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7755286 ↗

Other Light, Surgical, Fiberoptic devices