Skip to content

LITe DECOMPRESSION 48089999Recalls & reported problems

Stryker Corporation · LIGHT CABLE · Class 2

All models of LITe DECOMPRESSION (4)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613327064926), received between May 10, 2016 and Jul 5, 2016 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Device Emits Odor” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Device Emits Odor2
  • Adverse Event Without Identified Device or Use Problem1
  • Charred1
  • Material Deformation1
  • Sparking1
  • Thermal Decomposition of Device1

Patient problems

  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 5, 2016Thermal Decomposition of Device; Device Emits Odor; Material Deformation; Material Deformation

    IT WAS REPORTED THAT; NOTICED SMELL COMING FROM LIGHT SOURCE. WHEN I PULLED OUT LIGHT CABLE NOTICE IT WAS BURNED. I MEDIATED DISCONNECTED AND INSERTED NEW LIGHT CABLE IN NEW LIGHT SOURCE UNIT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5769473 ↗
  • MalfunctionMay 10, 2016Charred; Device Emits Odor; Sparking; Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT; SMELL WAS COMING OUT OF LIGHT SOURCE. WE SWITCHED TO NEW A SOURCE. NOTICE END OF LIGHT CABLE WAS CHARRED. WHEN WE SWITCHED LIGHT UNITS THE SMELL DISSIPATED. FEW MIN LATER, SURGEON PULLED OUT LIGHT CABLE FROM LIGHT DECOMPRESSION TUBE IMMEDIATELY. SAID IT SPARKED. UPDATE (B)(6) 2016: SURGEON SAID HE SAW A SPARK INSIDE TUBE WHICH WAS IN PATIENT. USED REPLACEMENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5645359 ↗

FDA clearance and classification

Cleared through 510(k) K122637 on Nov 1, 2012 (decision: Substantially Equivalent). Submitted by Stryker Spine as “Lite Decompression System- Light Cable Model 48089999”. The FDA review took 64 days. FDA record ↗

The FDA regulates product code FST (light, surgical, fiberoptic) under 21 CFR 878.4580 as a Class II device. 21 CFR 878.4580 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the LITe DECOMPRESSION 48089999

Has the LITe DECOMPRESSION 48089999 been recalled?

No FDA recall record in our data names this exact device identifier (DI 07613327064926), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the LITe DECOMPRESSION 48089999?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between May 10, 2016 and Jul 5, 2016. In those reports, the most common specific device problems are “Device Emits Odor” (2), “Charred” (1) and “Material Deformation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the LITe DECOMPRESSION 48089999?

None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the LITe DECOMPRESSION 48089999 FDA approved?

It was cleared through 510(k) K122637 on Nov 1, 2012. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the LITe DECOMPRESSION 48089999, and is it still sold?

The FDA's device database lists Stryker Corporation as the company responsible for it (brand LITe DECOMPRESSION, model 48089999). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other LIGHT, SURGICAL, FIBEROPTIC devices

All LIGHT, SURGICAL, FIBEROPTIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page