PhotonSaber Y Fin TipRecalls & reported problems
Invuity, Inc. · The Invuity PhotonSaber Y with Yankauer is intended to deliver illumination from a high intensity light source and to aspirate from the surgical site. · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00816728020455), received between Dec 20, 2021 and Dec 20, 2021 — classified as 1 malfunction.
The most frequently reported specific device problem is “Overheating of Device” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Overheating of Device1
Patient problems
- Partial thickness (Second Degree) Burn1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 20, 2021Overheating of Device
INVUITY PHOTONSABER Y - HANDHELD ILLUMINATOR WITH ASPIRATION DEVICE WAS CONNECTED PRIOR TO THE START OF THE PROCEDURE. PRODUCT WAS HOLSTERED CORRECTLY, BUT THE PATIENT SUSTAINED A 1 CM BLISTER/BURN TO THE LEFT INNER THIGH. THE AREA THAT HEATED UP TO CAUSE BLISTER IS WHERE THE LIGHT CORD CONNECTS TO THE LIGHTED SUCTION. ANOTHER PHOTONSABER WAS SWAPPED OUT AND USED, SO THE PHOTONSABER THAT CAUSED THE BLISTER WAS NEVER USED DURING THE PROCEDURE. WITH THE DEFECTIVE PHOTONSABER, STAFF WERE UNABLE TO REPLICATE THE CONDITIONS THAT LED TO THE HEATING UP OF THE CONNECTION AT THE BASE OF THE HANDLE TO THE LIGHT CORD. THE LIGHT CORD WAS ALSO PULLED FROM THE PROCEDURE AND CHECKED FOR ANY MALFUNCTION. THE LIGHT CORD WAS NOT FOUND TO HAVE ANY ISSUES. FDA SAFETY REPORT ID# (B)(4).
Patient problems: Partial thickness (Second Degree) Burn
MAUDE report 13059815 ↗
FDA clearance and classification
The FDA regulates product code JOL (catheter and tip, suction) under 21 CFR 880.6740 as a Class II device. 21 CFR 880.6740 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the PhotonSaber Y Fin Tip
Has the PhotonSaber Y Fin Tip been recalled?
No FDA recall record in our data names this exact device identifier (DI 00816728020455), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the PhotonSaber Y Fin Tip?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Dec 20, 2021. That report lists device problem “Overheating of Device” (1); patient problem “Partial thickness (Second Degree) Burn” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the PhotonSaber Y Fin Tip?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the PhotonSaber Y Fin Tip FDA approved?
Devices of this type (product code JOL) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Invuity, Inc.
Who makes the PhotonSaber Y Fin Tip, and is it still sold?
The FDA's device database lists Invuity, Inc. as the company responsible for it (brand PhotonSaber Y, model Fin Tip). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Catheter And Tip, Suction devices
- J-VAC 2160Ethicon Inc. · CATHETER AND TIP, SUCTION152 reports
- Wayne G56537Cook Incorporated · CATHETER AND TIP, SUCTION2 recalls33 reports
- Bard® Suction Connector EN0066000C. R. Bard, Inc. · CATHETER AND TIP, SUCTION1 recall6 reports
- ARROW IPN037088Teleflex Incorporated · CATHETER AND TIP, SUCTION1 recall0 reports
- MEDLINE DYND41478Medline Industries, Inc. · CATHETER AND TIP, SUCTION1 recall0 reports
- MEDLINE DYND41476Medline Industries, Inc. · CATHETER AND TIP, SUCTION1 recall0 reports
- ARROW IPN039282Teleflex Incorporated · CATHETER AND TIP, SUCTION1 recall0 reports
- MEDLINE DYND41474Medline Industries, Inc. · CATHETER AND TIP, SUCTION1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page