Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00816776020162), received between Mar 2, 2022 and Mar 2, 2022 — classified as 1 malfunction.
The most frequently reported specific device problem is “Missing Test Results” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Missing Test Results1
- Output Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 2, 2022Output Problem; Missing Test Results; Output Problem; Missing Test Results
THE CUSTOMER REPORTS THAT THE LAST COMPOUNDS IN THE LAST SAMPLE OF A BATCH SOMETIMES DOES NOT SHOW THE RESULTS PROPERLY WHEN USING AUTO REFRESH DUE TO A RACE CONDITION IN THE CODE. THE NEOLYNX. EXE PROGRAM RUNS AFTER EACH SAMPLE FINISHES ACQUIRING AND PROCESSES EACH INJECTION. THE CUSTOMER STARTS THEIR RUN, ANALYSES THE RESULTS AS ACQUIRED BY OPENING THE BROWSER & CHECKING INTENSITIES, THEN LEAVES THE BROWSER OPEN WITH AUTO-REFRESH ENABLED. WITH THE CUSTOMER'S BROWSER SET TO AUTO-REFRESH, NEOLYNX SENDS THE RESULTS TO THE BROWSER AS IT PROCESSES. A VULNERABILITY IN THE SOFTWARE MAY CAUSE NEOLYNX. EXE TO TELL THE BROWSER SOFTWARE TO REFRESH ITS DISPLAY BEFORE ALL THE DATA HAS FINISHED WRITING TO THE DISK. WHEN THIS HAPPENS, THE LAST RESULT IS INCOMPLETE IN THE BROWSER DISPLAY (BUT NOT IN THE RESULT FILE - NRF). THE SPECTRUM WINDOW IS NOT COMPLETELY LOADED, LOOKS BLANK AND THE FINAL CHANNELS OF THE LAST FUNCTION CAN BE MISSING. AT THE END OF THE RUN, THE CUSTOMER CHECKS FOR ERROR MESSAGES, REVIEWS THE "FAIL" FLAGGED SAMPLES ONLY AND PUSHES THE DATA LOADED TO LIMS. THE BROWSER ONLY SENDS THE DATA THAT IT HAS LOADED, SO THE ELECTRONIC FILE TO LIMS WILL NOT SHOW THE LAST DATA TRANSITIONS
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13655992 ↗
Other Mass Spectrometer, Clinical Use devices
- Citrine Triple Quad™ MS/MS 5050803AB Sciex (Manufacturing) · Mass Spectrometer, Clinical Use1 recall0 reports
- AB SCIEX Triple Quad™ 4500MD 5061724AB Sciex (Manufacturing) · Mass Spectrometer, Clinical Use1 recall0 reports
- Citrine QTRAP MS/MS 5050807AB Sciex (Manufacturing) · Mass Spectrometer, Clinical Use1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.