Citrine QTRAP MS/MS 5050807Recalls & reported problems

AB Sciex (Manufacturing) · The Citrine system is a liquid chromatography-tandem mass spectrometry (LC-MS/MS) systems intended to identify inorganic or organic compounds in human specimens by ionizing the compound under investigation and separating the resulting ions by means of an electrical field according to their mass. · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Aug 30, 2023Class IIOpen, Classified

    In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.

    Z-0039-2024 · Root cause: Component design/selection · Recall details · FDA record ↗