Microvention, Inc. · MicroPlex HyperSoft Helical · Class 2 · Implant

All models of V-Trak (75)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00816777020604), received between Oct 10, 2018 and Mar 7, 2026 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Device Markings/Labelling Problem1
  • Difficult or Delayed Positioning1
  • Difficult to Remove1
  • Separation Failure1
  • Stretched1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 7, 2026Difficult or Delayed Positioning

    IT WAS REPORTED THAT DURING A PROCEDURE THE COIL DID NOT RELEASE. A SECOND CONTROLLER OF A DIFFERENT LOT WAS USED; HOWEVER, THE SAME FAILURE WAS EXPERIENCED. IT WAS REPORTED THAT STANDARD TECHNICAL RECOMMENDATIONS WERE FOLLOWED, INCLUDING MOISTENING THE GOLD TIP OF THE COIL AND ROTATING THE CONTROLLER TO FACILITATE RELEASE. DESPITE THESE EFFORTS, THE AUDIBLE RELEASE SIGNAL WAS NOT HEARD ON ANY ATTEMPT, AND THE INDICATOR LIGHT CHANGED FROM GREEN TO ORANGE WITHOUT CONFIRMING DEPLOYMENT. IT WAS REPORTED THAT UNDER FLUOROSCOPIC GUIDANCE, THE COIL WAS CONFIRMED NOT TO HAVE RELEASED. THE PHYSICIAN PROCEEDED TO REMOVE THE SYSTEM, AND THROUGH COUNTER-TRACTION, THE COIL DETACHED WHILE REMAINING WITHIN THE MICROCATHETER. THE COIL WAS THEN POSITIONED WITHIN THE ANEURYSM USING ADVANCEMENT MANEUVERS WITH THE PUSHER, WHERE IT WAS ULTIMATELY IMPLANTED. THERE WAS NO PATIENT INJURY. IT IS INDICATED THAT THE PATIENT WAS IN GOOD CLINICAL CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24540365 ↗
  • MalfunctionMar 27, 2021Device Markings/Labelling Problem; Device Markings/Labelling Problem

    IT WAS REPORTED THAT THE COIL EXIT MARKER LOCATED ON THE PUSHER OUTSIDE OF THE PATIENT WAS NOT IN THE CORRECT LOCATION. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11579925 ↗
  • MalfunctionNov 14, 2019Stretched; Detachment of Device or Device Component; Stretched; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING COIL EMBOLIZATION OF AN ANEURYSM, THE COIL STRETCHED AND DETACHED DURING REPOSITIONING. THE IMPLANT WAS REMOVED IN ITS ENTIRETY ALONG WITH THE MICROCATHETER. THERE WAS NO REPORTED PATIENT INJURY OR INTERVENTION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9325402 ↗
  • InjuryOct 10, 2018Difficult to Remove; Separation Failure; Detachment of Device or Device Component; Difficult to Remove; Separation Failure; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE EMBOLIZATION COIL DID NOT DETACH. DURING THE ATTEMPT TO REMOVE THE COIL, IT STRETCHED AND DETACHED. THE COIL WAS REMOVED SUCCESSFULLY WITH A SNARE DEVICE. THERE WAS NO REPORTED PATIENT INJURY AS A RESULT OF THE ISSUE. THE PATIENT WAS REPORTED TO BE FINE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7954391 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →