V-Trak 100204HS-V-A2Recalls & reported problems
Microvention, Inc. · MicroPlex HyperSoft Helical · Class 2 · Implant
All models of V-Trak (75)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00816777020604), received between Oct 10, 2018 and Mar 7, 2026 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Device Markings/Labelling Problem1
- Difficult or Delayed Positioning1
- Difficult to Remove1
- Separation Failure1
- Stretched1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 7, 2026Difficult or Delayed Positioning
IT WAS REPORTED THAT DURING A PROCEDURE THE COIL DID NOT RELEASE. A SECOND CONTROLLER OF A DIFFERENT LOT WAS USED; HOWEVER, THE SAME FAILURE WAS EXPERIENCED. IT WAS REPORTED THAT STANDARD TECHNICAL RECOMMENDATIONS WERE FOLLOWED, INCLUDING MOISTENING THE GOLD TIP OF THE COIL AND ROTATING THE CONTROLLER TO FACILITATE RELEASE. DESPITE THESE EFFORTS, THE AUDIBLE RELEASE SIGNAL WAS NOT HEARD ON ANY ATTEMPT, AND THE INDICATOR LIGHT CHANGED FROM GREEN TO ORANGE WITHOUT CONFIRMING DEPLOYMENT. IT WAS REPORTED THAT UNDER FLUOROSCOPIC GUIDANCE, THE COIL WAS CONFIRMED NOT TO HAVE RELEASED. THE PHYSICIAN PROCEEDED TO REMOVE THE SYSTEM, AND THROUGH COUNTER-TRACTION, THE COIL DETACHED WHILE REMAINING WITHIN THE MICROCATHETER. THE COIL WAS THEN POSITIONED WITHIN THE ANEURYSM USING ADVANCEMENT MANEUVERS WITH THE PUSHER, WHERE IT WAS ULTIMATELY IMPLANTED. THERE WAS NO PATIENT INJURY. IT IS INDICATED THAT THE PATIENT WAS IN GOOD CLINICAL CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24540365 ↗MalfunctionMar 27, 2021Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT THE COIL EXIT MARKER LOCATED ON THE PUSHER OUTSIDE OF THE PATIENT WAS NOT IN THE CORRECT LOCATION. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11579925 ↗MalfunctionNov 14, 2019Stretched; Detachment of Device or Device Component; Stretched; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING COIL EMBOLIZATION OF AN ANEURYSM, THE COIL STRETCHED AND DETACHED DURING REPOSITIONING. THE IMPLANT WAS REMOVED IN ITS ENTIRETY ALONG WITH THE MICROCATHETER. THERE WAS NO REPORTED PATIENT INJURY OR INTERVENTION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9325402 ↗InjuryOct 10, 2018Difficult to Remove; Separation Failure; Detachment of Device or Device Component; Difficult to Remove; Separation Failure; Detachment of Device or Device Component
IT WAS REPORTED THAT THE EMBOLIZATION COIL DID NOT DETACH. DURING THE ATTEMPT TO REMOVE THE COIL, IT STRETCHED AND DETACHED. THE COIL WAS REMOVED SUCCESSFULLY WITH A SNARE DEVICE. THERE WAS NO REPORTED PATIENT INJURY AS A RESULT OF THE ISSUE. THE PATIENT WAS REPORTED TO BE FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7954391 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page