V-Trak 8410-0306-A2Recalls & reported problems
Microvention, Inc. · HyperSoft 3D Advanced · Class 2 · Implant
All models of V-Trak (75)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00816777021267), received between May 5, 2025 and May 5, 2025 — classified as 1 injury.
The most frequently reported specific device problem is “Unraveled Material” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Unraveled Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryMay 5, 2025Unraveled Material; Unraveled Material
IT WAS REPORTED: DURING TREATMENT OF ACUTELY RUPTURED ACOMM ANEURYSM, A SCEPTER XC WAS PLACED IN RIGHT M2 BRANCH AND A HEADWAY 17 MICROCATHETER WAS PLACED INTO THE BILOBED ANEURYSM. SUCCESSFUL DELIVERY OF 3 TERUMO NEURO COILS WAS PERFORMED WITH BALLOON INFLATION WITHOUT INCIDENT. THE FOURTH COIL WAS ATTEMPTED TO PLACE WITH REPOSITIONING OF THE COIL INTO THE EXISTING COIL MASS 3-4 TIMES. THE PHYSICIAN REALIZED THE COIL HAD STRETCHED AND DECIDED TO TAKE A SNARE OVER THE HEADWAY 17 MICROCATHETER AND WAS ABLE TO SUCCESSFULLY RETRIEVE THE STRETCHED COIL OUT OF THE PATIENT. THE COIL WAS NEVER DETACHED. IT HAD STRETCHED AND WAS STILL PARTIALLY IN THE ANEURYSM SAC UPON RETRIEVAL. THE INDWELLING COIL MASS WAS UNAFFECTED BY THE REMOVAL OF THE STRETCHED COIL. NO FURTHER INTERVENTION WAS REQUIRED. THE PATIENT WAS ADMITTED TO THE HOSPITAL WITH SUBARACHNOID HEMORRHAGE (SAH); IT WAS PRE-EXISTING AND NOT CAUSED BY THE PROCEDURE. THEY ARE UNABLE TO TELL CONDITION / OUTCOME CURRENTLY AND NO UPDATE ON PATIENT CONDITION IS AVAILABLE AS OF REPORT DATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21965064 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page