HydroFrame 10 Advanced 100612HFRM-V-A2Recalls & reported problems

Microvention, Inc. · V-Trak · Class 2 · Implant

All models of HydroFrame 10 Advanced (22)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00816777023544), received between Mar 11, 2022 and Jul 22, 2025 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Material Separation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Material Separation1
  • Migration1
  • Premature Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Stroke/CVA1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 22, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT THE COIL BROKE DURING INSERTION AND HAD TO BE RECOVERED FROM THE PATIENT. THEY HAD ALREADY PLACED SOME COILS INTO THE ANEURYSM. THE AFFECTED COIL COULD BE ADVANCED THROUGH THE MICROCATHETER WITHOUT ANY ISSUES AND WAS DEPLOYED INTO THE ANEURYSM FOR APPROXIMATELY THE FIRST 2 MILLIMETERS. AFTER THAT, THE MICROCATHETER WAS PUSHED OUT OF THE ANEURYSM, AND THE COIL COULD NO LONGER BE MANIPULATED VIA THE PUSHER WIRE. THE COIL APPEARS TO HAVE BROKEN INSIDE THE MICROCATHETER ¿ BY ADVANCING A GUIDEWIRE THROUGH THE MICROCATHETER, THE DISTAL PART BEYOND THE FRACTURE SITE WAS CLAMPED. THIS ALLOWED THE DETACHED COIL TO BE RETRIEVED FROM THE PATIENT. THE PATIENT IS DOING WELL ¿ THE RECOVERY CAUSED NO HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22568831 ↗
  • MalfunctionMay 17, 2024Premature Separation; Premature Separation

    IT WAS REPORTED THAT AN EMBOLIZATION COIL IMPLANT DETACHED PREMATURELY INSIDE THE MICROCATHETER. THE IMPLANT WAS REMOVED IN ITS ENTIRETY ALONG WITH THE MICROCATHETER. THERE WAS NO INJURY OR INTERVENTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19346905 ↗
  • MalfunctionMar 11, 2022Adverse Event Without Identified Device or Use Problem; Migration; Adverse Event Without Identified Device or Use Problem; Migration

    IT WAS REPORTED THAT DURING AN EMBOLIZATION TREATMENT OF A RIGHT MEDIA ANEURYSM ON (B)(6) 2021 FOUR HYDROFRAME-10 AV COILS WERE DEPLOYED WITHOUT INCIDENT. ON (B)(6) 2021 IMAGING REVEALED SOME SMALL MICRO-INFARCTS. THE PATIENT WAS REPORTED IN GOOD CONDITION WITHOUT PERMANENT NEUROLOGICAL DAMAGE. NO SECONDARY INTERVENTION WAS TAKEN AND NO SEQUELAE WAS REPORTED.

    Patient problems: Stroke/CVA

    MAUDE report 13741184 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →