0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00816777023766), received between Oct 11, 2022 and Jun 30, 2026 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Unraveled Material” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unraveled Material2
  • Detachment of Device or Device Component1
  • Premature Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 30, 2026Unraveled Material

    AS REPORTED: THE COIL STRETCHED AND UNWOUND AND NEEDED TO BE TACKED DOWN WITH A STENT. DURING RECAPTURE AND TRYING TO REPOSITION MULTIPLE TIMES THE COIL BECAME COMPROMISED OR DAMAGED AND WHEN PULLED BACK INTO MICROCATHETER STRETCHED. ATTEMPTS WERE MADE TO REMOVE BUT WOULD HAVE CAUSED DISRUPTION OF COIL MASS SO DECISION WAS MADE TO LEAVE. FOLLOW UP SHOWED NO IMMEDIATE ISSUE OR COMPLICATION. PHYSICIAN WILL DUE FOLLOW UP IN 6 MONTHS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25673262 ↗
  • InjuryMay 19, 2026Unraveled Material; Premature Separation

    AS REPORTED: EMBOLIZATION COIL UNRAVELED AND BROKE APART, IS RETAINED INSIDE THE LEFT ICA OF PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 25219921 ↗
  • MalfunctionOct 11, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN EMBOLIZATION PROCEDURE, TWO (2) CONSECUTIVE COILS DID NOT FIT THROUGH THE MICROCATHETER. BOTH COILS STARTED TO CATCH IN THE LUMEN AND DETACHED IN THE CATHETER PRIOR TO USE OF THE CONTROLLER. THE COILS WERE REMOVED IN ONE PIECE, AND ADDITIONAL COILS WERE USED FOR THE REMAINDER OF THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY OR ISSUES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15580746 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →