Microvention, Inc. · MicroPlex HyperSoft Helical · Class 2 · Implant

All models of V-Trak (75)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00816777024114), received between Oct 25, 2018 and Jul 1, 2026 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Component Missing” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Component Missing2
  • Stretched2
  • Inadequacy of Device Shape and/or Size1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Thrombus1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 1, 2026Component Missing

    IT WAS REPORTED THAT THE CONCERNED EMBOLIZATION COIL WAS BEING USED IN CONJUNCTION WITH A COIL-ASSIST STENT TO TREAT A PATIENT WITH A LEFT PCOM ANEURYSM USING "JAILING" TECHNIQUE. AFTER PLACING THE STENT (REF. MFR REPORT#: 2032493-2026-08049), THE PHYSICIAN STARTED COILING THE ANEURYSM WITH THE CONCERNED COIL AND REPORTEDLY NOTED THAT THE 3CM PROXIMAL RADIOPAQUE MARKER WAS MISSING FROM THE COIL. THE PHYSICIAN USED FLUOROSCOPY TO ASSESS WHETHER THE COIL WAS INSIDE THE ANEURYSMAL SAC IN ORDER TO DETACH IT, AS THIS CAN BE MISLEADING AT CERTAIN ANGLES. THE PHYSICIAN COULD SEE WHEN THE COIL WAS ON THE TIP OF THE MICROCATHETER FOR THE DETACHMENT ZONE. THE COIL WAS DETACHED AND IMPLANTED IN THE ANEURYSM. THE PATIENT WAS DOING VERY WELL AFTER THE PROCEDURE AND WAS ABLE TO MOVE BOTH ARMS AND LEGS; AND ABLE TO ANSWER USUAL QUESTIONS. THIS IS EVENT REPORT 2 OF 2.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25681827 ↗
  • MalfunctionJun 23, 2026Component Missing

    AS REPORTED: WHEN THE COIL WAS REMOVED FROM THE MICROCATHETER, DR. NOTICED THAT THERE WAS NO RADIOPAQUE MARKING ON THE PROXIMAL END OF THE COIL. THE COIL WAS REMOVED AS A PRECAUTION BUT WAS DISPOSED OF DIRECTLY IN THE WASTE BIN. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25591473 ↗
  • InjuryJan 11, 2022Stretched; Stretched

    IT WAS REPORTED THAT DURING AN EMBOLIZATION TREATMENT, THE PUSHER OF AN EMBOLIZATION COIL IMPLANT STRETCHED DURING POSITIONING. A STENT WAS USED TO AFFIX THE PUSHER TO THE VESSEL WALL BUT THE STRETCHED PORTION EXTENDED INTO THE CAROTID ARTERY. A STENT WAS USED TO REMOVE THE REMAINING PORTION OF THE COIL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13232794 ↗
  • MalfunctionAug 18, 2021Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    IT WAS REPORTED THAT AN EMBOLIZATION COIL IMPLANT KEPT COMING OUT OF THE ANEURYSM NECK DURING PLACEMENT. THE COIL WAS REMOVED FROM THE PATIENT WHERE IT WAS OBSERVED THAT THE COIL WAS THE INCORRECT SIZE. THERE WAS NO REPORTED PATIENT INJURY OR INTERVENTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12340476 ↗
  • InjuryOct 25, 2018Stretched; Stretched

    IT WAS REPORTED THAT DURING REPOSITIONING OF THE EMBOLIZATION COIL IN AN ANEURYSM, THE COIL STRETCHED AND THEN COULD NOT BE WITHDRAWN INTO THE MICROCATHETER. THE COIL WAS PUSHED FORWARD UNTIL IT WAS FULLY WITHIN THE ANEURYSM AND WAS THEN INTENTIONALLY DETACHED MANUALLY WTIHOUT THE CONTROLLER. FOLLOW-UP ANGIOGRAPHY DEMONSTRATED THROMBUS AT THE NECK OF THE ANEURYSM AND IN THE VESSEL. IV INTEGRILIN WAS ADMINISTERED AS TREATMENT. THE PATIENT WAS REPORTED TO BE ALRIGHT AND HAD A GOOD OUTCOME.

    Patient problems: Thrombus

    MAUDE report 8007117 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →