AZUR 45-780202-LRecalls & reported problems

Microvention, Inc. · Azur CX Detachable · Class 2 · Implant

All models of AZUR (14)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00816777024367), received between Aug 24, 2017 and Mar 25, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Detachment Of Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment Of Device Component2
  • Kinked2
  • Premature Separation2
  • Detachment of Device or Device Component1

Patient problems

  • No Consequences Or Impact To Patient3
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 25, 2026Premature Separation

    IT WAS REPORTED THAT AN EMBOLIZATION COIL IMPLANT WOULD NOT DETACH WITH TWO DIFFERENT DETACHMENT CONTROLLERS. IT WAS THEN NOTICED THAT THE COIL HAD UNEXPECTEDLY DETACHED. THE PATIENT WAS REPORTED TO BE "FINE AND IT WAS A SUCCESSFUL PROCEDURE." THE COIL WAS IN APPROPRIATE POSITION WHEN DETACHMENT WAS ATTEMPTED SO IT WAS LEFT IN PLACE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24684185 ↗
  • MalfunctionMar 22, 2026Premature Separation

    IT WAS REPORTED THAT THE PHYSICIAN POSITIONED THE COIL AND, UPON ATTEMPTING TO DEPLOY IT, OBSERVED A RED LIGHT ON THE DETACHMENT HANDLE. THE PHYSICIAN WIPED THE HANDLE AND ATTEMPTED DEPLOYMENT AGAIN, WITH THE SAME RED-LIGHT RESPONSE. A SECOND DETACHMENT HANDLE WAS USED, BUT THE SAME RED LIGHT OCCURRED. WHEN THE PHYSICIAN ATTEMPTED TO REMOVE THE COIL, IT WAS DISCOVERED THAT THE COIL HAD IN FACT DETACHED, ALTHOUGH NO GREEN LIGHT OR DEPLOYMENT LIGHT SEQUENCE WAS OBSERVED ON THE HANDLE. IT WAS REPORTED THAT THE EVENT DID NOT AFFECT THE PATIENT. THE COIL WAS DEPLOYED IN THE INTENDED LOCATION, AND THERE WAS NO BLOOD LOSS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24658911 ↗
  • MalfunctionApr 30, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING RE-POSITIONING OF THE COIL IN THE INTENDED TREATMENT AREA, THE COIL DETACHED WITHOUT USE OF THE V-GRIP CONTROLLER. THE COIL REMAINS IMPLANTED IN THE PATIENT. THERE WAS NO REPORTED PATIENT INJURY OR INTERVENTION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8567530 ↗
  • MalfunctionAug 24, 2017Detachment Of Device Component; Kinked; Detachment Of Device Component; Kinked

    IT WAS REPORTED THAT DURING POSITIONING OF AN EMBOLIZATION COIL, THE COIL UNEXPECTEDLY DETACHED AND KINKED IN THE MICROCATHETER. THE COIL WAS REMOVED IN ITS ENTIRETY WITH THE MICROCATHETER. THERE WAS NO REPORTED INTERVENTION OR PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6819653 ↗
  • MalfunctionAug 24, 2017Detachment Of Device Component; Kinked; Detachment Of Device Component; Kinked

    IT WAS REPORTED THAT DURING POSITIONING OF AN EMBOLIZATION COIL, THE COIL UNEXPECTEDLY DETACHED AND KINKED IN THE MICROCATHETER. THE COIL WAS REMOVED IN ITS ENTIRETY WITH THE MICROCATHETER. THERE WAS NO REPORTED INTERVENTION OR PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6819651 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →