HydroFill Helical 100615HFIL-V-A2Recalls & reported problems
Microvention, Inc. · V-Trak · Class 2 · Implant
All models of HydroFill Helical (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00816777024404), received between Apr 23, 2021 and Apr 22, 2025 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Difficult to Remove1
- Migration or Expulsion of Device1
- Premature Separation1
- Stretched1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Foreign Body In Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 22, 2025Premature Separation; Premature Separation
IT WAS REPORTED THAT THE PHYSICIAN WAS TREATING AN ANEURYSM ON THE RIGHT MEDIA BIFURCATION. SHE WANTED TO COIL THE ANEURYSM USING A REMODELING TECHNIQUE. AFTER PLACING SEVERAL COILS, A HYDROFILL WAS CHOSEN. WHILE REPOSITIONING THE COIL PREMATURELY DETACHED. THE COIL WAS EXCHANGED AND THE EXAMINATION CONTINUED. AFTER SUCCESSFUL COILING THE IMPLANTATION OF A STENT WAS PLANNED AND CHOSEN. THE STENT PREMATURELY DETACHED IN THE MICROCATHETER. THE STENT WAS EXCHANGED AND THE EXAMINATION CONTINUED WITH GOOD RESULT. THE PATIENT IS DOING WELL. NO INJURY TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21887227 ↗InjuryMay 8, 2024Migration or Expulsion of Device; Detachment of Device or Device Component; Migration or Expulsion of Device; Detachment of Device or Device Component
IT WAS REPORTED THAT THE PHYSICIAN PUT THE COÏL IN ANEURYSM (ACA). IT WAS THE SECOND COÏL. THE PHYSICIAN DECIDED TO REPOSITION THIS COÏL AFTER COMPLETE DEPLOYMENT. WHEN IT WAS RETRIEVED, THE COÏL BROKE AND AFTER SEVERAL ATTEMPTS TO RETRIEVE ALL OF THE COÏL, A HALF PART OF COÏL WAS LEFT IN THE VESSEL, CAROTID ARTERY. THE PHYSICIAN WAS THEN OBLIGED TO PUT TWO CAROTID STENTS INSIDE THE VESSEL TO MAINTAIN THE BROKEN COÏL. ADDITIONAL INFORMATION WAS RECEIVED AND IT WAS REPORTED THAT, "WHEN THE COIL WAS DETACHED USING THE DETACHMENT SYSTEM, THE SECTION WAS COMPLETED, BUT THE PUSHER CATHETER REMAINED ATTACHED. AFTER SEVERAL ATTEMPTS TO RETRIEVE THE COIL AND PUSHER WITH RETRIEVER STENTS, IT ONLY ¿DESCENDED¿ TO THE LEVEL OF THE COMMON CAROTID ARTERY. IT WAS NECESSARY TO STENT THE CAROTID ARTERY WITH 2 CAROTID STENTS IN ORDER TO PLATE THE PUSHER AND STENT ALONG THE CAROTID WALL."
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 19274207 ↗InjuryApr 23, 2021Difficult to Remove; Stretched; Detachment of Device or Device Component; Difficult to Remove; Stretched; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A COIL EMBOLIZATION, THE COIL STRETCHED AND PREMATURELY DETACHED IN THE VESSEL. THE COIL WAS RETRIEVED WITH AN ENDOVASCULAR SNARE. THE PATIENT WAS REPORTED TO BE FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11712773 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page