HydroFrame 10 Advanced 7110-0410-A2Recalls & reported problems

Microvention, Inc. · V-Trak · Class 2 · Implant

All models of HydroFrame 10 Advanced (22)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00816777025234), received between Dec 26, 2018 and May 24, 2024 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Device Markings/Labelling Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Device Markings/Labelling Problem1
  • Device Misassembled During Manufacturing /Shipping1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Hydrocephalus1
  • No Patient Involvement1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 24, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED THROUGH THE RAGE CLINICAL STUDY: BELOW EVENT WAS REVIEWED BY INDEPENDENT REVIEWER (CEA) AND WAS DETERMINED TO BE POSSIBLY RELATED TO STUDY DEVICE AND UNLIKELY RELATED TO STUDY PROCEDURE. EVENT DESCRIPTION: SUBJECT HAD MILD SUPRATENTORIAL VENTRICULOMEGALY WITH FLUCTUATION IN CLINICAL STATUS. IT WAS CLARIFIED THE SUBJECT HAD 2 LUMBAR DRAINS INSERTED ONE ON (B)(6) 2022 AND ANOTHER ON (B)(6) 2022. NO SIGNIFICANT NEUROLOGICAL CHANGES, LUMBAR DRAIN REMOVED. EVENT TERM: HYDROCEPHALUS. EVENT START DATE: 09/MAR/2022. POST OPERATIVE DAYS: 33 DAYS. AE OUTCOME : RESOLVED WITHOUT SEQUELAE - DATE OF RESOLUTION WAS (B)(6) 2022 . DEVICES INVOLVED: HYDROFRAME LOT 2102181WF 4 X 10 ¿ MODEL 7110-0410-A2. HYDROFRAME LOT 0000091908 3 X 6 MODEL 7110-0306-A2. HYDROSOFT 3D LOT 0000139563 3 X 6 UNKNOWN MODEL . HYDROSOFT LOT 0000091013 2.5 X 6 MODEL 7210-0256-A2. HYDROSOFT LOT 0000109762 2 X 3 MODEL 7210-0203-A2. AS IT CANNOT BE DETERMINED WHICH COIL DEVICE MAY HAVE IMPACTED THE EVENT, HYDROFRAME 4X10, MODEL 7110-0410-A2 / LOT 2102181WF WAS CHOSEN AS A DEFAULT PRIMARY DEVICE. THE OTHER COILS USED IN THE PROCEDURE WILL BE LISTED AS CONCOMITANT DEVICES.

    Patient problems: Hydrocephalus

    MAUDE report 19397935 ↗
  • InjuryMay 5, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED THROUGH THE RAGE CLINICAL STUDY, COIL COMPACTION WAS OBSERVED 16 MONTHS POST TREATMENT. RETREATMENT WAS PERFORMED TO TREAT THE COIL COMPACTION. RESOLVED WITHOUT SEQUELAE ON (B)(6) 2023. NO PATIENT HARM OR INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16882626 ↗
  • MalfunctionFeb 18, 2022Device Markings/Labelling Problem; Device Markings/Labelling Problem

    IT WAS REPORTED THAT DURING PREPARATION, IT WAS NOTICED THAT AN EMBOLIZATION IMPLANT COIL'S PACKAGE AND BOX HAD MISMATCHED LABELS. THE COIL WAS PLACED ASIDE AND THE PROCEDURE WAS COMPLETED WITH ANOTHER COIL. THE PRODUCT DID NOT CONTACT THE PATIENT. NINE OTHER COILS USED DURING THE CASE WERE CHECKED AND THEIR PACKAGING WAS CORRECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13563786 ↗
  • MalfunctionDec 26, 2018Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    IT WAS REPORTED THAT AFTER THE COIL HAD BEEN REMOVED FROM THE STERILE POUCH AND THE SCRUB TECH WAS GETTING READY TO PREP THE COIL, THE COMPANY REPRESENTATIVE NOTICED THAT THE COIL HOOP DID NOT LOOK "RIGHT" AND CHECKED THE DEVICE PACKAGE LABELS, WHICH INDICATED AN AZUR CX 18 PERIPHERAL COIL WAS INSIDE THE HYDROFRAME 10 NEURO COIL BOX. THE COIL WAS NOT USED IN THE PATIENT.

    Patient problems: No Patient Involvement

    MAUDE report 8196849 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →