Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00816777025364), received between May 18, 2017 and May 23, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Premature Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 23, 2025Premature Separation; Premature Separation

    IT WAS REPORTED THAT THE COIL DEPLOYED HALF IN THE BRAIN, HALF IN THE CATHETER. NO ADDITIONAL INFORMATION AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22081774 ↗
  • MalfunctionMay 18, 2017Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT TREATMENT WAS PERFORMED ON A 3MM ANTERIOR COMMUNICATING ARTERY ANEURYSM VIA THE RIGHT INTERNAL CAROTID ARTERY. AN EMBOLIZATION COIL WAS ADVANCED QUICKLY THROUGH THE MICROCATHETER WITHOUT USE OF FLUOROSCOPY AND A SLIGHT "POP" WAS NOTED. UNDER FLUOROSCOPY, THE COIL WAS FOUND TO HAVE DEPLOYED IN THE ANEURYSM WITHOUT USE OF THE CONTROLLER. AS THE PUSHER WAS RETRACTED, THE COIL IMPLANT AND DISTAL PUSHER DID NOT MOVE. THE PROXIMAL HALF OF THE PUSHER WAS REMOVED FROM THE MICROCATHETER AND IT WAS NOTED TO BE FRACTURED. THE NON-DETACHED COIL IMPLANT AND FRACTURED DISTAL END OF THE PUSHER WAS CAREFULLY REMOVED ALONG WITH THE MICROCATHETER. ALL COMPONENTS OF THE COIL WERE REMOVED FROM THE PATIENT WITHOUT INCIDENT. THERE WAS NO REPORTED INTERVENTION OR PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6577541 ↗