V-Trak 7110-0254-3D-A2Recalls & reported problems

Microvention, Inc. · HydroSoft 3D · Class 2 · Implant

All models of V-Trak (75)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00816777025425), received between Jul 25, 2018 and Apr 1, 2025 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Device Dislodged or Dislocated1
  • Migration or Expulsion of Device1
  • Stretched1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1
  • Paraplegia1
  • Stroke/CVA1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 1, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    AS REPORTED: HYDROSOFT COIL MIGRATED AFTER SUCCESSFUL BALLOON ASSISTED COILING INTERVENTION FROM THE SAME LEFT A1 ANEURYSM. THE A2 DID NOT HAVE FLOW FOR SEVERAL MINUTES AND THE TOTAL TIME UNTIL FULL FLOW WAS RESTORED WAS A TOTAL OF 1 HOUR; THIS RESULTED IN A STROKE. DR. DID NOT PERFORM STENT ASSISTED COILING BECAUSE THE CORRECT LENGTH DEVICE WAS NOT ON THEIR SHELF AND HE DID NOT WANT TO COVER THE ANTERIOR CHORIDAL. DR. TRIED TO RETRIEVE THE COIL WITH A THROMBUS RETRIEVAL DEVICE (2 PASSES) AND SMALL MICROSNARE ALONG WITH ASPIRATION. THE DECISION WAS MADE TO UTILIZE A STENT 2.5X23 (172020) TO KEEP THE COIL ALONGSIDE A1 SEGMENT VESSEL WALL AND RESTORE FLOW. A1 EXPERIENCED VESSEL SPASM, BUT NO EVIDENCE OF ANY DAMAGE CONFIRMED WITH IN SUITE CTA. THE PATIENT AFTER WAKING UP COULD NOT MOVE HIS RIGHT LEG OR ARM AND WAS APHASIC. SYMPTOMS STARTED TO IMPROVE. PATIENT IS CURRENTLY TALKING AND ABLE TO MOVE HIS RIGHT HAND, THE RIGHT LEG IS STILL NOT MOVING.

    Patient problems: Paraplegia; Stroke/CVA

    MAUDE report 21748453 ↗
  • MalfunctionDec 9, 2020Stretched; Detachment of Device or Device Component; Stretched; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING STENT-ASSISTED COIL EMBOLIZATION OF AN ANTERIOR COMMUNICATING ARTERY (ACOM) ANEURYSM, THE EMBOLIZATION COIL STRETCHED IN THE ANEURYSM AND THEN DETACHED. THE MICROCATHETER WAS REMOVED AND A STRETCHED SEGMENT OF THE COIL REMAINS IN THE PARENT ARTERY. NO ADDITIONAL INTERVENTION WAS PERFORMED. THERE WAS NO REPORTED PATIENT INJURY. THE PATIENT WAS REPORTED TO BE DOING WELL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10980793 ↗
  • MalfunctionJul 25, 2018Migration or Expulsion of Device; Migration or Expulsion of Device

    IT WAS REPORTED THAT DURING AN EMBOLIZATION TREATMENT FOR A RUPTURED ANEURYSM, THE COIL WAS IMPLANTED. CT IMAGES TAKEN 3 DAYS LATER SHOWED THE COIL HAD DISLODGED AND MIGRATED DOWNSTREAM TO THE RIGHT DISTAL INFERIOR DIVISION. NO INTERVENTION WAS PERFORMED FOR THE MIGRATED COIL. THE INITIAL TREATMENT SITE WAS RE-COILED WITH STENT ASSISTANCE THE DAY AFTER THE CT WAS PERFORMED. THE PATIENT WAS DISCHARGED AND IS REPORTED TO BE ALERT AND ORIENTED, AND THERE ARE NO NEUROLOGIC DEFICITS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7721281 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →