HydroSoft 3D 7110-0310-3D-A2Recalls & reported problems
Microvention, Inc. · V-Trak · Class 2 · Implant
All models of HydroSoft 3D (26)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00816777025487), received between May 6, 2019 and Jan 29, 2026 — classified as 5 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component3
- Stretched2
- Unraveled Material2
- Adverse Event Without Identified Device or Use Problem1
- Contamination /Decontamination Problem1
- Device-Device Incompatibility1
- Premature Separation1
- Separation Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient2
- Diplopia/Double Vision1
- Memory Loss/Impairment1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryJan 29, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED FOR A PATIENT ENROLLED IN THE CAR REGISTRY STUDY FOR A RUPTURED ANEURYSM TREATMENT OF ANTERIOR COMMUNICATING ARTERY THAT THE PATIENT EXPERIENCES INTERMITTENT DIPLOPIA, ESPECIALLY WHEN TIRED. EVENT ONSET DATE: AUGUST 31, 2025 . AE RELATEDNESS: RELATIONSHIP TO STUDY DEVICE: POSSIBLY RELATED . RELATIONSHIP TO INDEX ENDOVASCULAR PROCEDURE: POSSIBLY RELATED. RELATIONSHIP TO USE OF ANCILLARY DEVICE: NOT RELATED. RELATIONSHIP TO STUDY DISEASE: PROBABLY RELATED. RELATIONSHIP TO CONCURRENT DISEASE: NOT RELATED. RELATIONSHIP TO DUAL ANTIPLATELET THERAPY: NOT RELATED. POST OPERATIVE DAYS: 39 WAS THE AE CAUSED BY OR ASSOCIATED WITH A DEVICE DEFICIENCY? NO OUTCOME: RESOLVED. ADDITIONAL INFORMATION REPORTED THAT THE PATIENT REPORTING DIFFICULTY REMEMBERING NEW INFORMATION. EVENT ONSET DATE: AUGUST 31, 2025 . AE RELATEDNESS: RELATIONSHIP TO STUDY DEVICE: POSSIBLY RELATED, RELATIONSHIP TO INDEX ENDOVASCULAR PROCEDURE: POSSIBLY RELATED, RELATIONSHIP TO USE OF ANCILLARY DEVICE: NOT RELATED, RELATIONSHIP TO STUDY DISEASE: POSSIBLY RELATED, RELATIONSHIP TO CONCURRENT DISEASE: POSSIBLY RELATED, RELATIONSHIP TO DUAL ANTIPLATELET THERAPY: NOT RELATED. POST OPERATIVE DAYS: 39 WAS THE AE CA
Patient problems: Diplopia/Double Vision; Memory Loss/Impairment
MAUDE report 24218183 ↗InjuryJan 17, 2025Unraveled Material; Unraveled Material
NO ADDITIONAL INCIDENT INFORMATION WAS PROVIDED, DEVICE WAS DISCARDED. PLEASE SEE H6 & H11.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21186033 ↗MalfunctionDec 9, 2024Contamination /Decontamination Problem; Contamination /Decontamination Problem
ADDITIONAL INFORMATION RECEIVED THAT PROVIDED AN IMAGE AND WAS REVIEWED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20887300 ↗InjuryJul 1, 2022Premature Separation; Premature Separation
IT WAS REPORTED THE PHYSICIAN WAS TREATING A BASILAR TIP ANEURYSM WITH Y STENT AND MULTIPLE COILS. INITIAL COILS INSERTED AND IMPLANTED WITHOUT ISSUE. THE 3X10 HYDROSOFT 3D COIL PARTIALLY DEPLOYED INTO ANEURYSM AND THE PHYSICIAN TRIED TO RE-SHEATH AND RE-DEPLOY, BUT THE COIL PREMATURELY DETACHED INSIDE THE CATHETER. THE COIL WAS SNARED AND REMOVED FROM THE PATIENT. THE PROCEDURE CONTINUED AND SUCCESSFULLY COMPLETED USING A FEW MORE COILS OF DIFFERENT BRAND. THERE WAS NO HARM OR INJURY TO THE PATIENT, WHO REPORTEDLY WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14894324 ↗InjuryMar 3, 2022Stretched; Unraveled Material; Detachment of Device or Device Component; Stretched; Unraveled Material; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING AN ANEURYSM EMBOLIZATION PROCEDURE, THE COIL WAS DEPLOYED AS THE THIRD AND FINAL COIL. MULTIPLE DEPLOYMENT AND RE-SHEATHING ATTEMPTS WERE MADE WHEN THE PHYSICIAN NOTED THE COIL FILE STRETCHED AND DETACHED. ULTIMATELY IT SEEMED THE COIL DETACHED WITH A LONG TETHER REMAINING PROXIMAL TO THE END OF THE COIL. THE PROXIMAL 1-2CM OF COIL WERE LEFT OUT OF THE ANEURYSM WITH AN ADDITIONAL PORTION LEFT PROXIMAL TO THE LVIS JR. STENT, WHICH HAD ALREADY BEEN DEPLOYED, JAILING THE MICROCATHETER BEING USED TO COIL. AS WELL, THIS REMAINING TETHER (OF AROUND 5-10CM) WAS LEFT PROXIMAL TO THE LVIS JR. AND ULTIMATELY TACKED DOWN WITH AN ATLAS STENT, PROXIMAL TO THE ORIGINAL COIL/STENT CONSTRUCT. MULTIPLE ATTEMPTS AT CLEARING THE MICROCATHETER WITH A GUIDEWIRE FAILED TO DISLODGE THE TETHER, WHICH WAS STILL IN THE MICROCATHETER. A SNARE WAS ALSO UTILIZED AROUND THE MICROCATHETER IN ORDER TO SNARE THE TETHER OR COIL TO REMOVE IT, HOWEVER, THE SNARE WOULD NOT NAVIGATE THE ANATOMY AND COULD NOT REACH THE SITE. ULTIMATELY IT WAS DECIDED TO TACK DOWN THE TAIL WITH AN ATLAS STENT. THE PATIENT IS DOING WELL AT END AND POST OF PROCEDURE. NO FURTHER TREATMENT PERFORMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13662184 ↗InjuryMar 20, 2020Stretched; Separation Failure; Detachment of Device or Device Component; Stretched; Separation Failure; Detachment of Device or Device Component
IT WAS REPORTED THAT TREATMENT WAS PERFORMED ON A PARTIALLY THROMBOSED P-COM ANEURYSM. AFTER PLACEMENT OF THE EMBOLIZATION COIL, THE COIL COULD NOT BE DETACHED WITH THREE DIFFERENT DETACHMENT CONTROLLERS OR A 9-VOLT BATTERY. DURING THE ATTEMPT TO REMOVE THE COIL, THE COIL STRETCHED. THE COIL WAS BROKEN OFF AND TACKED AGAINST THE VESSEL WALL WITH THREE (3) STENTS. APPROXIMATELY 1 CM OF IMPLANT COIL EXTENDED INTO THE P-COM AND ANOTHER 6 CM OF STRETCHED COILS BACKED DOWN INTO THE CAVERNOUS SEGMENT. THERE WAS NO REPORTED PATIENT INJURY OR CLINICAL SEQUELA.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9862350 ↗MalfunctionMay 6, 2019Detachment of Device or Device Component; Device-Device Incompatibility; Detachment of Device or Device Component; Device-Device Incompatibility
IT WAS REPORTED THAT FRICTION WAS ENCOUNTERED DURING THE ADVANCEMENT OF THE 3RD COIL IN A SUBARACHNOID HEMORRHAGE ANTERIOR COMMUNICATING ARTERY ANEURYSM, AND THE COIL GOT STUCK INSIDE THE CATHETER. DURING REMOVAL, THE COIL DETACHED IN THE CATHETER. THERE WAS NO REPORTED PATIENT INJURY OR INTERVENTION. THE PATIENT WAS REPORTED TO BE DOING WELL.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8584671 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page