HydroSoft 3D 100206HFRM-V-3D-A2Recalls & reported problems
Microvention, Inc. · V-Trak · Class 2 · Implant
All models of HydroSoft 3D (26)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00816777026514), received between Feb 6, 2020 and Dec 27, 2024 — classified as 2 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Device-Device Incompatibility2
- Adverse Event Without Identified Device or Use Problem1
- Material Separation1
- Premature Separation1
- Stretched1
- Unraveled Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient2
- Muscle Weakness/Atrophy1
- Ruptured Aneurysm1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 27, 2024Material Separation; Stretched; Premature Separation; Material Separation; Stretched; Premature Separation
AS REPORTED: MCA ANEURYSM TREATMENT, USUAL COIL USAGE STARTED. DURING PREPARATION OF COIL, THEY REALISED SOME WIRE DEBRIS ON THE HYPOTUBE. STARTED THE ACCESS INTO THE MICROCATHETER AND THE DEVICE WAS DETACHED FROM THE PUSHER. REMOVED ALL PARTS AND COLLECTED. PATIENT TREATED WITH OTHER COILS SUCCESSFULLY. PATIENT IS IN GOOD CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21028320 ↗InjuryDec 6, 2024Unraveled Material; Unraveled Material
InjuryMar 18, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PROCEDURE WAS FOR TREATMENT OF AN ACUTELY RUPTURED A2 SEGMENT ANEURYSM. THE EMBOLIZATION COIL WAS BEING USED AS THE SECOND COIL WHEN IT UNINTENTIONALLY DETACHED FROM THE PUSHER WIRE. A BALLOON WAS THEN INTRODUCED AND INFLATED TO PREVENT COIL PROLAPSING INTO PARENT VESSEL AND THE DESCRIBED COIL WAS PUSHED INTO THE ANEURYSM. ONCE THE COIL WAS SUCCESSFULLY PLACED WITHIN THE ANEURYSM, THE OPERATOR PULLED THE PUSHER WIRE BACK BUT THE PROXIMAL END OF THE COIL ALSO PULLED BACK. THE DETACHMENT CONTROLLER WAS THEN ATTACHED AND THE COIL WAS DETACHED SUCCESSFULLY. A SMALL AMOUNT OF EXTRAVASATION/RUPTURE OF THE ANEURYSM WAS NOTED, WHICH WAS QUICKLY TREATED WITH FURTHER COILS BEING PLACED. PATIENT INITIALLY HAD RIGHT SIDED WEAKNESS IN UPPER AND LOWER LIMB AND REPORTEDLY REGAINED LIMB POWER OVER FOLLOWING DAYS.
Patient problems: Muscle Weakness/Atrophy; Ruptured Aneurysm
MAUDE report 13805520 ↗MalfunctionMar 19, 2020Detachment of Device or Device Component; Device-Device Incompatibility; Detachment of Device or Device Component; Device-Device Incompatibility
IT WAS REPORTED THAT BALLOON-ASSISTED COILING WAS PERFORMED FOR AN MIDDLE CEREBRAL ARTERY (MCA) BIFURCATION ANEURYSM. DURING PLACEMENT OF THE THIRD COIL, RESISTANCE WAS ENCOUNTERED AND DURING THE ATTEMPT TO REMOVE THE COIL, IT DETACHED IN THE MICROCATHETER. BOTH SEGMENTS OF THE COIL WERE REMOVED TOGETHER WITH THE MICROCATHETER. THERE WAS NO REPORTED PATIENT INJURY OR ADDITIONAL INTERVENTION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9857954 ↗MalfunctionFeb 6, 2020Detachment of Device or Device Component; Device-Device Incompatibility; Detachment of Device or Device Component; Device-Device Incompatibility
IT WAS REPORTED THAT COIL EMBOLIZATION WAS PERFORMED FOR AN INTERNAL CAROTID ARTERY ANEURYSM. WHILE ADVANCING THE THIRD COIL, RESISTANCE WAS ENCOUNTERED AND DURING WITHDRAWAL, THE COIL DETACHED INSIDE THE MICROCATHETER. BOTH SEGMENTS OF THE COIL WERE REMOVED IN THEIR ENTIRETY TOGETHER WITH THE MICROCATHETER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9679408 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page