Skip to content

HydroSoft 3D 100254HFRM-V-3D-A2Recalls & reported problems

Microvention, Inc. · V-Trak · Class 2 · Implant

All models of HydroSoft 3D (26)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00816777026538), received between Nov 20, 2023 and Sep 26, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1
  • Premature Separation1
  • Separation Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjurySep 26, 2024Device Dislodged or Dislocated; Separation Problem; Device Dislodged or Dislocated; Separation Problem

    AS REPORTED: FOURTH COIL WAS DEPLOYED INTO RUPTURED ANEURYSM, COULD NOT GET ENTIRE LENGTH OF THE COIL INTO THE ANEURYSM, APPROXIMATELY 1CM OF COIL REMAINED PROTRUDING ON EACH OF 3 ATTEMPTS. INCREASED FRICTION FELT MANIPULATING COIL IN ADVANCED STRAIGHT MICROCATHETER, SO ATTEMPTED TO REMOVE COIL. HOWEVER, THE PUSHER WIRE OF THE COIL SNAPPED PROXIMALLY. DISTAL FRAGMENT OF COIL AND PUSHER WIRE WERE REMOVED WITH A SNARE AND THE REMAINING TAIL OF COIL WAS STENTED IN PLACE TO SUCCESSFULLY COMPLETE THE PROCEDURE. SURGICAL INTERVENTION IN FORM OF STENT BEING DEPLOYED, HOWEVER NO ADVERSE EFFECTS TO PATIENT NOTED AS A RESULT OF THIS PROCEDURE. NO CHANGE TO MRS PRE / POST PROCEDURE. PROCEDURE PERFORMED: ACOM ANEURYSM EMBOLISATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20313861 ↗
  • MalfunctionNov 20, 2023Premature Separation; Premature Separation

    SEE H10.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18178809 ↗

Other DEVICE, NEUROVASCULAR EMBOLIZATION devices

All DEVICE, NEUROVASCULAR EMBOLIZATION devices →