VSP System Final Surgical SplintRecalls & reported problems

3d Systems, Inc. · A physical template / guide that conforms to patients’ anatomy (Bony and / or Occlusal) to allow manipulation of free floating bony anatomy. · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00816847020541), received between Aug 12, 2020 and Feb 4, 2026 — classified as 6 injuries.

The most frequently reported specific device problem is “Device-Device Incompatibility” (1 report).

1 other recall cover devices cleared under the same FDA submission (K120956, K133907, K192192); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Device-Device Incompatibility1
  • Patient Device Interaction Problem1
  • Patient-Device Incompatibility1
  • Product Quality Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Joint Dislocation1
  • Malunion of Bone1
  • No Code Available1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 4, 2026Adverse Event Without Identified Device or Use Problem

    DURING A MAXILLOFACIAL PROCEDURE INVOLVING VSP SYSTEM DEVICES, THE SURGEON REPORTED THAT THE PRE-DRILLED HOLES FOR THE CUSTOM PLATE DID NOT PROPERLY ALIGN, LEADING TO ADDITIONAL HOLES BEING DRILLED IN THE PATIENT'S MAXILLA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24260418 ↗
  • InjuryFeb 16, 2023Device-Device Incompatibility; Device-Device Incompatibility

    WHILE PERFORMING CORRECTIVE JAW SURGERY UTILIZING VSP SYSTEMS' LEFORT DRILLING GUIDE AND SURGICAL SPLINTS, THE SURGEON REPORTED THE LEFORT DRILLING GUIDE AND SURGICAL SPLINT DID NOT FIT AS ANTICIPATED. ADDITIONALLY, THE SURGEON REPORTED THAT THE SPLINTS NEEDED TO BE ALTERED IN THE OPERATING ROOM TO ACHIEVE A SATISFACTORY FIT. THESE ISSUES CONTRIBUTED TO A SURGICAL DELAY OF 120+ MINUTES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16391090 ↗
  • InjuryFeb 15, 2023Patient Device Interaction Problem; Patient Device Interaction Problem

    DURING AN ORTHOGNATHIC SURGICAL PROCEDURE, THE SURGEON HAD DIFFICULTY USING THE INTERMEDIATE AND FINAL SURGICAL SPLINT TO SET THE PATIENT'S FINAL OCCLUSION. THIS RESULTED IN A REVISION SURGERY THAT TOOK PLACE ON (B)(6) 2022 SO THE SURGEON COULD FIX THE PATIENT'S OCCLUSION.

    Patient problems: Malunion of Bone

    MAUDE report 16380910 ↗
  • InjuryNov 18, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DURING A COMBINATION CASE INVOLVING VIRTUAL SURGICAL PLANNING (VSP) AND STRYKER TMJ-CONCEPTS, THE SURGEON REPORTED A MALOCCLUSION AFTER PLACEMENT OF THE FINAL SURGICAL SPLINT. THE SURGEON REPORTED A 90-MINUTE DELAY IN SURGERY AS THEY ATTEMPTED TO ADJUST THE PATIENT'S OCCLUSION INTRAOPERATIVELY.

    Patient problems: Joint Dislocation

    MAUDE report 15826424 ↗
  • InjuryMar 24, 2022Product Quality Problem; Product Quality Problem

    THE VSP SYSTEM FINAL SURGICAL SPLINT DID NOT FIT THE SURGEON'S PLANNED FINAL POSITION OF THE PATIENT'S MANDIBLE DURING A VSP ORTHOGNATHIC TMJ SURGERY. THE SURGICAL DELAY WAS REPORTED TO BE 1-2 HOURS FOR ACHIEVING THE SPLINT TO FIT THE PATIENT'S FINAL OCCLUSION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13891031 ↗
  • InjuryAug 12, 2020Patient-Device Incompatibility; Patient-Device Incompatibility

    THE SURGEON REPORTED THE VSP SYSTEM FINAL SURGICAL SPLINT AND FINAL PALATAL SPLINT DID NOT MATCH THE SURGEON'S PLANNED FINAL POSITION OF THE PATIENT'S MAXILLA FOR THE MAXILLARY/CONDYLECTOMY ORTHOGNATHIC SURGERY.

    Patient problems: No Code Available

    MAUDE report 10396979 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K120956, K133907, K192192. One submission often covers many models, so check the product and lot details in each record.

  • Z-1510-2022 · May 17, 2022 · Open, Classified
    3D Systems VSP Reconstruction, Model: VSPR-301 Product: Scapula Marking Guide (Anatomically Shaped), Rx Only, Non-Sterile

Other System, Image Processing, Radiological devices

All System, Image Processing, Radiological devices →