VSP System Mandible Marking GuideRecalls & reported problems
3d Systems, Inc. · A physical template / guide that is positioned on patients’ anatomy (Bony and / or Occulsal), and allows for precise location of predetermined osteotomy planes and holes. · Class 2
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00816847020824), received between Jun 30, 2022 and Mar 17, 2026 — classified as 6 injuries, 1 malfunction.
The most frequently reported specific device problem is “Malposition of Device” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K120956, K133907, K192192); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem5
- Malposition of Device2
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Failure of Implant2
- Nerve Damage1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryMar 17, 2026Adverse Event Without Identified Device or Use Problem
WHILE PERFORMING THE VERTICAL OSTEOTOMY OF THE BSSO, THE RESIDENT ASSUMED THE CUTTING SLOT ENDED AT THE INFERIOR BORDER AND CONTINUED CUTTING BASED OFF THAT. CONFUSION ALSO AROSE WHEN LOOKING AT THE NERVE TO DISTAL MANDIBLE MEASUREMENTS IN THE CASE REPORT AS THE RESIDENT BELIEVED THERE SHOULD'VE BEEN TWO MEASUREMENTS - ONE AT THE ANTERIOR CUT OF THE WEDGE AND ANOTHER AT THE POSTERIOR. THESE COMBINED RESULTED IN THE PATIENTS NERVES BEING SEVERED BILATERALLY.
Patient problems: Nerve Damage
MAUDE report 24617726 ↗InjuryFeb 6, 2026Adverse Event Without Identified Device or Use Problem
SURGEON REPORTED DIFFICULTIES DURING A MAXILLOFACIAL PROCEDURE INVOLVING VSP SYSTEM DEVICES. SURGICAL COMPLICATIONS RELATED TO THE FIT OF THE CUSTOM PLATES RESULTED IN THE SCHEDULING OF A REVISION SURGERY TO ADDRESS CONCERNS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24279956 ↗MalfunctionJun 18, 2024Malposition of Device; Malposition of Device
FOLLOWING A MAXILLOFACIAL PROCEDURE UTILIZING VSP SYSTEMS PRODUCTS, IT WAS REPORTED THAT THE SURGEON WAS UNHAPPY WITH THE SURGICAL OUTCOME AND INTENDS TO PERFORM A REVISION SURGERY. IN PARTICULAR, THE SURGEON NOTED DIFFICULTY UTILIZING THE PROVIDED MANDIBLE CUTTING GUIDES.
Patient problems: Failure of Implant
MAUDE report 19559935 ↗InjuryMay 23, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
FOLLOWING A MAXILLOFACIAL PROCEDURE UTILIZING VSP SYSTEMS PRODUCTS, IT WAS REPORTED THAT THE SURGEON HAD OPERATIVE DIFFICULTIES UTILIZING THE PROVIDED 3D SYSTEMS MANDIBLE MARKING GUIDE AND SUBSEQUENT DIFFICULTIES IN THE PLACEMENT OF THE MANDIBLE PROSTHESIS. THIS RESULTED IN AN OPERATIVE DELAY OF MORE THAN TWO HOURS AND IT HAS BEEN REPORTED THAT A REVISION SURGERY IS BEING PLANNED.
Patient problems: Failure of Implant
MAUDE report 19383290 ↗InjuryApr 24, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
FOLLOWING A MAXILLOFACIAL SURGERY, IT WAS REPORTED THAT THE SURGEON HAD DIFFICULTY FITTING A MANDIBLE PROSTHESIS TO THE PATIENT USING A 3D SYSTEMS MANDIBLE MARKING GUIDE, WHICH CAUSED A DELAY IN THE OPERATION OF APPROXIMATELY 4 HOURS. THE PROCEDURE WAS COMPLETED WITHOUT ADDITIONAL PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19175138 ↗InjuryNov 15, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DURING A CONDYLECTOMY PROCEDURE, A SURGEON REPORTED DISCOVERING A MISALIGNMENT BETWEEN THE PREDICTED SCREW HOLE LOCATIONS OF A 3D SYSTEM'S DEVICE AND THE PATIENT'S ANATOMY. ADDITIONAL HOLES WERE DRILLED IN THE PATIENT'S MANDIBLE TO ENSURE PROPER IMPLANT PLACEMENT. POST-PROCEDURE INVESTIGATION REVEALED NO IDENTIFIABLE PRODUCT OR PROCESS DEFICIENCIES THAT WOULD ACCOUNT FOR THIS MIS-ALIGNMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18141891 ↗InjuryJun 30, 2022Malposition of Device; Malposition of Device
DURING A VSP ORTHOGNATHICS AND TMJ SURGERY, THERE WERE ISSUES WITH THE FIT OF THE MANDIBULAR PROSTHESIS. THE SURGEON DRILLED THE PREDICTIVE HOLES INDICATED BY THE MANDIBLE MARKING GUIDE AND LATER FOUND THAT THE MANDIBULAR PROSTHESIS WOULD NOT MEET THE FOSSA ONCE IT WAS SCREWED INTO THE MANDIBLE. AS A RESULT, THE SURGEON DID NOT USE THE PRE-DRILLED HOLES AND PREFERRED A DIFFERENT POSITIONING OF THE MANDIBLE PROSTHESIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14881088 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K120956, K133907, K192192. One submission often covers many models, so check the product and lot details in each record.
- Z-1510-2022 · May 17, 2022 · Open, Classified3D Systems VSP Reconstruction, Model: VSPR-301 Product: Scapula Marking Guide (Anatomically Shaped), Rx Only, Non-Sterile
Other System, Image Processing, Radiological devices
- Interventional Workspot Interventional Workspot - 1Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological13 recalls24 reports
- Synapse Radiology PACS Synapse Radiology PACS Software LicenseFujifilm Healthcare Americas Corporation · System, image processing, radiological9 recalls38 reports
- CENTRICITY UNIVERSAL VIEWER ZFP CLIENT 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER 7.0Ge Healthcare · System, image processing, radiological6 recalls0 reports
- IntelliSpace Cardiovascular 8Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological2 recalls132 reports
- CA1000 2.0 SP 14Ge Healthcare · System, image processing, radiological4 recalls0 reports
- VMA 2Ortho Kinematics, Inc. · System, Image Processing, Radiological4 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.