Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00816988011491), received between Oct 25, 2017 and Apr 3, 2019 — classified as 3 injuries.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break2
- Detachment of Device or Device Component1
- Difficult to Remove1
- Improper or Incorrect Procedure or Method1
Patient problems
- Device Embedded In Tissue or Plaque1
- Foreign Body In Patient1
- No Information1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryApr 3, 2019Break; Difficult to Remove; Detachment of Device or Device Component
IT WAS REPORTED DURING A LEFT ANKLE FRACTURE PROCEDURE, THE PROXIMAL END OF THE FIB30130L FIBULA NAIL BROKE OFF DURING INSERTION AND WAS LEFT INSIDE OF THE FIBULA. THE REP STATED THE FRACTURE WAS SUCCESSFULLY REDUCED AND FIXATED WITH A DISTAL FIBULA PLATE AND SCREWS. THE SEATED DEPTH OF THE IMPLANT WAS 130MM. A 3.22MM WAS USED TO PREP THE BONE. ADDITIONAL INFORMATION HAS BEEN REQUESTED. ADDITIONAL INFORMATION RECEIVED ON (B)(6) 2019: THE REP STATED THAT THE SURGEON ATTEMPTED TO REMOVE THE BROKEN FRAGMENT, BUT NO AVAIL. THE REP CONFIRMED THE REMAINING PORTION OF THE NAIL WAS RETRIEVED AND WILL BE RETURNING FOR EVALUATION. THE BONE WAS PREPPED BY USING A 6.2 DISTAL REAMER AND 3.2 PROXIMAL REAMER. THE REP STATED THAT THE SURGEON HAD TO SWITCH TO THEIR BACK UP PLAN TO MAKE A LARGER INCISION SO THEY COULD PLATE THE FRACTURE. A SECOND FIB30130L WAS NOT BROUGHT IN. THE CASE WAS COMPLETED BY USING AN ANKLE FRACTURE PLATE.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 8479859 ↗InjuryMay 24, 2018Improper or Incorrect Procedure or Method
IT WAS REPORTED THAT DURING A DISTAL FIBULA FRACTURE PROCEDURE, THE SURGEON DRILLED WITH THE 6.2 MM AND THEN THE 3.2 MM DRILLS FROM THE FIBULOCK DISPOSABLE KIT, PLACED CANNULA, PROCEEDED TO PLACE THE NAIL BY HAND, BUT REALIZED HE SHOULD HAVE DRILLED A BIT FURTHER. HE PULLED THE NAIL BACK OUT OF THE FIBULA WITH EASE, BUT THE TALONS STAYED IN THE FIBULA. HE WAS ABLE TO RE-DRILL AND PLACE A NEW NAIL, WHICH TURNED OUT GREAT, BUT THE EXTRA TALONS STAYED IN THE PATIENT. NO ATTEMPTS WERE MADE TO REMOVE THE TALON/TOP PIECE OF THE DEVICE.
Patient problems: Foreign Body In Patient
MAUDE report 7540109 ↗InjuryOct 25, 2017Break
IT WAS REPORTED DURING A CASE THAT THE 3.0X130MM FIBULA NAIL, LEFT BROKE OFF IN THE FIBULA AS THE NAIL WAS BEING DRIVEN IN. THIS HAPPENED PRIOR TO HAVING TO DEPLOY THE TALONS. THE SURGEON REAMED TO PROPER DEPTH WITH BOTH THE 6.2 AND 3.2 DRILL. THE SURGEON HAD TO CUT THE PROXIMAL PORTION OF THE NAIL TO GET THE NAIL IN. FOLLOW-UP INVESTIGATION: PER SALES REP THE CASE WAS COMPLETED BY CUTTING THE NAIL SHORT AND FINISHING THE CASE PER SURGICAL TECHNIQUE. NO FURTHER ISSUES REPORTED.
Patient problems: No Information
MAUDE report 6975638 ↗
Other Rod, Fixation, Intramedullary And Accessories devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page